Keysight N774xC Series Multiport Power Meter Programming Guide
37
Instrument Setup and Status
3
Command:
:STATus[n]:OPERation:ENABle
Syntax:
:STATus[n]:OPERation:ENABle<wsp><value>
Description:
Sets the bits in the Operation Slot Status Enable Mask (OSSEM) for slot
n
that enable the contents of the Operation Slot
Status Event Register (OSSER) for slot
n
to affect the OSESR.
Setting a bit in this register to 1 enables the corresponding bit in the OSSER for slot
n
to affect bit
n
of the OSESR.
Parameters:
The bit value for the OSSEM as a
16-bit unsigned integer
value (0 .. +65535)
Response:
none
Example:
:stat0:oper:enab 128
Command:
:STATus[n]:OPERation:ENABle?
Syntax:
:STATus[n]:OPERation:ENABle?
Description:
Returns the OSSEM of slot n
Parameters:
none
Response:
The bit value for the OSSEM as a
16-bit unsigned integer
value (0 .. +65535)
Example:
:stat0:oper:enab? -> +128
Command:
:STATus:PRESet
Syntax:
:STATus:PRESet
Description:
Presets all bits in all OPERation and QUEStionable status systems to 0.
Parameters:
none
Response:
none
Example:
:stat:pres
Command:
:STATus:QUEStionable[:EVENt]?
Syntax:
:STATus:QUEStionable[:EVENt]?
Description:
Returns the Questionable Status Event Summary Register (QSESR).