34
N7026A 150 MHz AC/DC High Sensitivity Current Probes User’s Guide
N7026A Probe
.
Figure 12
Complete Test Setup
11
Measure the current and record
V
rms
output of the probe as displayed on the DMM <2>.
12
Calculate gain as follows:
Gain =
V
rms
* scale factor
Gain should be between 0.99 and 1.01 (+/-1% gain accuracy).
13
Remove the test cable and AC current generator from the N7026A probe.