N7026A Probe
N7026A 150 MHz AC/DC High Sensitivity Current Probes User’s Guide
33
Figure 11
Setting up AC Current Generator
9
Set the DMM <2> to V
rms
mode.
10
As displayed in
Figure 12
, connect a BNC cable to the:
a
N1022B adapter’s output using the
NMD 3.5mm (M) to 3.5mm (F) adapter
and
3.5mm (M)
to BNC (F) adapter
.
b
DMM <2> using a
BNC (F)-to-banana adapter
.
This completes the test setup.