49
Keysight N7020A Power Rail Probe
User’s Guide
6 Performance Verification
Procedure
50
Performance Test Record
51
This chapter tests that the probe meets its specified input resistance, which is
50 k
Ω
±2%.
CAUTION
Electrostatic discharge (ESD) can quickly and imperceptibly damage or destroy high
performance probes, resulting in costly repairs. Always wear a wrist strap when handling
probe components and ensure that cables are discharged before being connected.
NOTE
Allow the probe to warm up for at least 20 minutes.
Table 9
Required Test Equipment
Test Equipment
Critical Specification
Model Number
Oscilloscope
Supported Oscilloscope
S-Series or 9000A series
Infiniium Oscilloscope
Digital Multimeter
2 wire resistance accuracy >
±
0.01%
34401A or equivalent
Adapter
BNC (f) to SMA (m) (In E2655C Kit)
E2655-83201
Содержание N7020A
Страница 1: ...Keysight N7020A Power Rail Probe User s Guide ...
Страница 22: ...22 N7020A User s Guide 1 General Information ...
Страница 32: ...32 N7020A User s Guide 2 Using the N7020A Figure 10 ESD Workstation ...
Страница 36: ...36 N7020A User s Guide 3 Calibrating ...
Страница 47: ...Performance Plots 5 N7020A User s Guide 47 Figure 17 Input Impedance Equivalent Model ...
Страница 48: ...48 N7020A User s Guide 5 Performance Plots ...
Страница 52: ...52 N7020A User s Guide 6 Performance Verification ...
Страница 54: ...54 N7020A User s Guide Index ...