This chapter explains requirements for using the probe, how to connect the probe
to the Device Under Test (DUT), and how to protect the oscilloscope and probe
from damage due to ESD.
CAUTION
Always wear an ESD wrist strap when working with active probes. Not doing so can result
in damage to the probe and to the oscilloscope’s input channel.
NOTE
If the N7020A input is connected to an S-series oscilloscope at vertical settings at or below
20 mV/div with more than 200 mV of applied offset with no signal on the probe input, the
probe output may be indeterminate. Upon zeroing the offset, increasing the vertical setting, or
connecting the probe input to a low impedance (power rail), the output will be correctly
restored.