Disassembly
2
Keysight RP7900 Series Advanced Service Guide
43
Disassembly Procedures
Removing the top chassis
1
First, remove all cables, wires, or power cord that are connected to the
instrument before performing any disassembly.
2
Remove the 7 screws (0515-1946) on the top chassis of the instrument.
3
Next, remove the 3 screws (0515-0430) at the rear panel of the instrument.