1
Troubleshooting
30
Keysight RP7900 Series Advanced Service Guide
315
Settings conflict error
A data element could not be programmed because of the present instrument state.
Refer to the
Operating and
Service Guide
316
Mass storage error
The mass storage memory has been exceeded.
317
Invalid format
An invalid data format was found in the command string.
320
Firmware update error
This may be due to the instrument hardware not being able to support the firmware
version.
Check the hardware revision or
firmware programming setup
324
Inconsistent arb settings
The arb settings are inconsistent; most likely a mismatch in the arb lengths.
Refer to the
Operating and
Service Guide
327
Initiated with no sense function enabled
A measurement has been initiated without specifying the
measurement(sense)function.
328
Too many measurement points
Too many measurement points have been specified.
331
Illegal parameter value
The parameter value is out of range or does not exist.
332
Master/slave error
An error has occurred in the master/slave configuration
Check the master/slave
connection or the hardware
333
Safety Disconnect error
An error has occurred in the SDS unit.
Check the SDS connection or the
hardware
Error
code Description
Potential remedy