64 Installation Note N5247-90113
Installation Procedure for the Upgrade
-
— Receiver Adjustment
— LFE Receiver Adjustment
— IF Response Adjustment (Options S93090xA/B, S93093A/B, or S93094A/B
Only)
— Noise Figure Adjustment (Option 029 with S93029A/B Only)
These adjustments are described in the PNA Service Guide and in the PNA
on-line HELP. A list of equipment required to perform these adjustments is also
found in the service guide.
To view this service guide information, click the Chapter 3 bookmark “Tests
and Adjustments” in the PDF Service Guide
1
.
After the specified adjustments have been performed, the analyzer should
operate and phase lock over its entire frequency range.
EEPROM Backup
The analyzer uses arrays of correction constants to enable the analyzer to
produce accurate, leveled source signals and receive clean test signals. These
constants are stored in non-volatile EEPROM memory and in flash memory
files.
The adjustments listed here generate new correction constants. The analyzer
must have a backup of this new data in case any of the data becomes
corrupted.
To store the backup data, perform these steps:
— Navigate to the EEPROM Backup Utility, located at:
—
Windows 7 --
C:\Program Files (x86)\Keysight\Network
Analyzer\Service\eebackup.exe
— Windows 10 --
C:\Program Files\Keysight\Network
Analyzer\Service\eebackup.exe
— Run the program.
— Click Backup EEPROM.
— Click Backup TSMB Mem.
— Click Backup Synth Mem. (Applies to Version 7 Synthesizers Only)
— Click Exit when the program has finished.
1. See
Содержание N5247-60113
Страница 3: ...3 ...
Страница 4: ...4 ...
Страница 66: ...66 Installation Note N5247 90113 Installation Procedure for the Upgrade ...
Страница 67: ...Installation Note N5247 90113 67 Installation Procedure for the Upgrade ...
Страница 68: ...68 Installation Note N5247 90113 Installation Procedure for the Upgrade ...
Страница 75: ...Installation Note Xxxxx xxxxx 3 ...