Installation Note N5247-90113 61
Installation Procedure for the Upgrade
Step 32. Enable Option 425
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must
not
be running.
— A keyboard must be connected to the network analyzer.
Option Enable Procedure
1.
Locate the email(s) from Keysight which contain license file attachments.
These emails are the result of
“Step 1. Obtain a Keyword and Verify the
.
Copy the license file(s) from the email(s) to a USB flash drive.
More than one license file may be copied to the USB flash drive.
3.
Insert the USB flash drive to the PNA-X’s USB drive slot. Within 5 seconds,
the PNA-X should display a small “New licenses installed” message.
Else, load the license key file(s), manually move your license file(s) to
C:\Program Files\Agilent\licensing
. It may take Keysight License
Manager an extra ~5 seconds to enable the licenses.
4.
Verifying and editing the license file:
For these steps, refer to the example in
a.
Verify your USB flash drive is connected to a PC.
b.
Open your license file using a text read/write program similar to
Notepad.
c.
If you have more than one licensed feature, delete the feature that is
not
required for this upgrade. (e.g., in this case N5242B-423 is the
correct upgrade. So, N5242B-422 is to be deleted from the text file.)
For this step, you will need a USB flash drive.
A single license file may contain more than one feature.
A single license file may contain more than one feature.
Содержание N5247-60113
Страница 3: ...3 ...
Страница 4: ...4 ...
Страница 66: ...66 Installation Note N5247 90113 Installation Procedure for the Upgrade ...
Страница 67: ...Installation Note N5247 90113 67 Installation Procedure for the Upgrade ...
Страница 68: ...68 Installation Note N5247 90113 Installation Procedure for the Upgrade ...
Страница 75: ...Installation Note Xxxxx xxxxx 3 ...