N4960A Getting Started Guide
14
Keysight N4960A Serial BERT 17 and 32 Gb/s Getting Started Guide
This instrument has autoranging line voltage input. Be sure the supply voltage is
within the specified range.
4.
In the MAIN menu position the arrow next to the Pat Gen-Jit
Menu label then press the softkey corresponding to the SEL
label.
5.
In the Pat Gen-Jit Menu position the arrow next to the Data
Output Menu label then press the softkey corresponding to the
SEL label.
6.
In the Data Output Menu position the arrow next to the Enable
label then press the softkey corresponding to the EDIT label.
7.
Use the rotary knob to set Enable to ON.
8.
Press the softkey corresponding to the SEL label to accept the
entry and return to the Data Output Menu.
9.
Press the softkey corresponding to the BACK label until the
MAIN menu is displayed.
10.
In the MAIN menu position the arrow next to the Err Det Menu
label then press the softkey corresponding to the SEL label.
11.
In the Err Det Menu position the arrow next to the AutoAlign
Menu label then press the softkey corresponding to the SEL
label.
12.
Position the arrow next to the Perform AutoAlign label then
press the softkey corresponding to the EXEC label. A small clock
appears in the display. When the clock disappears, ensure that
the Errors, Data Loss, and Sync Loss LEDs on the front panel of
the error detector are off.
13.
Press the softkey corresponding to the BACK label to return to
the Err Det Menu.
14.
In the Err Det Menu position the arrow next to the Accumulation
Menu label then press the softkey corresponding to the SEL
label.
15.
Position the arrow next to the Start Accum label then press the
softkey corresponding to the START label. The Run LED on the
front panel of the error detector should be lit.
16.
After 15 to 30 seconds, press the softkey corresponding to the
STOP label.
17.
Position the arrow next to the Results label then press the
softkey corresponding to the SEL label.
18.
Ensure that Errs reads 0.000e0.