N4960A Getting Started Guide
Keysight N4960A Serial BERT 17 and 32 Gb/s Getting Started Guide
13
1.5
Unpacking
Carefully remove the N4960A Serial BERT 17 and 32 Gb/s components
from the case in an ESD-safe environment.
1.6
Important Notes
Use ESD protection at all times when using the instrument.
Before connecting any cable to the instrument, discharge the cable
by shorting the center and outer connectors of the cable together to
ground momentarily.
Install the instrument on a flat surface away from heat sources.
Do not block the fans, or the exhaust vents on the rear and side
panels (3” min. clearance).
Use a 8 lbf-in (90 N-cm) torque wrench when attaching connectors.
1.7
Installation
1.
Plug the AC power cord into the N4960A-CJ0 or N4960A-CJ1
serial BERT controller rear panel power socket.
2.
Plug the other end of the AC power cord into a suitable wall
socket. (100-240 VAC, 50/60 Hz).
If this product is not used as specified, the protection provided by the equipment
could be impaired. This product must be used in a normal condition (in which all
means for protection are intact) only.
3.
Interconnect the equipment as shown in
Figure 2. Test Setup