Performance Verification

N2804/5A Differential Probes User’s Guide

33

12

On the oscilloscope, note the measured DC amplitude for the square wave input to 

the probe. For N2804A probes, divide this value by 100 and record the voltage 

below as the probe’s output (V

probe out

). For N2805A probes, divide the measured 

value by 50.
V

probe out

:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ____________ mV

13

Calculate the probe gain accuracy. V

cal

 was recorded in 

step 5

 and V

probe out

 was 

recorded in 

step 12

14

Verify that the probe gain accuracy is ±1% + scope gain accuracy. Record the test 

results as DC Differential Gain Accuracy in 

Table 10

 on page 34.

Bandwidth

15

Disable the 9500B’s output.

16

Connect the 50

W

 BNC feed through adapter to the N9500B’s 

CH1

 active head.

17

Attach the BNC (f)-to-banana post adapter to the BNC feed through adapter. 

18

Attach the differential probe input leads by clipping the alligator clamp to the BNC 

adapter banana post.

19

Enable the 9500B’s output.

20

Configure the oscilloscope to the following settings:
Amplitude Scale (Channel 1): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .500 mV/div

Time Scale: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2 ns/div

Acquisition mode: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Peak Detect

21

Center the trace on the oscilloscope.

22

On the 9500B, configure 

CH1

 to the following settings:

Waveform:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . sine wave

Load: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 M

W

Amplitude:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3V pk-pk

Frequency (N2804A):  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .300 MHz

Frequency (N2805A):  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .200 MHz

Probe Gain Accuracy (%)

V

probe out

V

cal

V

cal

----------------------------------------

100

×

=

Содержание N2804

Страница 1: ...Keysight N2804 5A Differential Probes User s Guide...

Страница 2: ...ware to U S government customers under its standard commer cial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www key sight com find sweula The...

Страница 3: ...obe Accessories 14 Probe Offset Adjustment 15 Safety Information 17 Specifications and Characteristics 20 N2804A Performance Plots 23 N2805A Performance Plots 27 Performance Verification 31 Performanc...

Страница 4: ...4 N2804 5A Differential Probes User s Guide Contents Index...

Страница 5: ...ce of 4 MW and low input capacitance of 4 pF to minimize circuit loading The probe comes with a pair of extension leads 30 cm long with a damping resistor built in to damp out the in band resonance an...

Страница 6: ...omplete or damaged notify your Keysight Technologies Sales Office Inspect the probe If there is mechanical damage or defect or if the probe does not operate properly or pass calibration tests notify y...

Страница 7: ...ing You should also avoid any mechanical shocks to this product in order to guarantee accurate performance and protection Figure 2 N2805A Product Labels Figure 3 N2804A Product Labels CAUTION Always w...

Страница 8: ...fix known defects and incorporate product enhancements To download the latest firmware go to www keysight com locate the Technical Support page and search by your oscilloscope model number NOTE The N2...

Страница 9: ...colored ring near the probe s channel connector and place an identical color ring near the probe head The channel identifications rings are shown in N2804A in Case on page 10 and N2805A in Case on pa...

Страница 10: ...th N2805A probes Replacements can be ordered with the N2816A replacement kit The quantity for each accessory in the kit is the same as listed in the table and originally provided with the probe For IE...

Страница 11: ...th Probe Order N2816A for Replacement Accessories Accessory Description Alligator Clips 1 red and 1 black Rating 600V 10A CAT III BW 300 MHz Pincer Clips 1 red and 1 black Rating 300V 1A CAT II BW 300...

Страница 12: ...th N2804A probes Replacements can be ordered with the N2817A replacement kit The quantity for each accessory in the kit is the same as listed in the table and originally provided with the probe For IE...

Страница 13: ...acement Accessories Accessory Description Alligator Clips 1 red and 1 black Rating 300V 10A CAT II BW 200 MHz Hook Clips 1 red and 1 black Rating 1000V 3A CAT III BW 200 MHz Pincer Clips 1 red and 1 b...

Страница 14: ...nsure that the probe s output connector is attached to the channel input of the oscilloscope and the oscilloscope is properly grounded CAUTION To protect against electrical shock use only the accessor...

Страница 15: ...n the oscilloscope and wait 20 minutes to allow the oscilloscope and probe to warm up 2 If the oscilloscope needs calibration perform a user calibration before the probe calibration 3 Using the probe...

Страница 16: ...e oscilloscope to Averaging mode x8 or higher to reduce oscilloscope noise 7 Set the vertical scale of the oscilloscope to 100 mV div for N2804A and 50 mV div for N2805A 8 Using the offset adjustment...

Страница 17: ...ety precautions Be aware that if you use this probe assembly in a manner not specified the protection this product provides may be impaired WARNING Observe Maximum Working Voltage To avoid injury do n...

Страница 18: ...ualified service personnel WARNING Connect and Disconnect Properly Connect the probe to the oscilloscope and connect the ground lead to earth ground before connecting the probe to the circuit under te...

Страница 19: ...st connect the protective earth terminal of the instrument to the protective conductor of the mains power cord The mains plug shall only be inserted in a socket outlet provided with a protective earth...

Страница 20: ...is an Industrial Scientific and Medical Group 1 Class A product ICES NMB 001 indicates product compliance with the Canadian Interference Causing Equipment Standard This product complies with the WEEE...

Страница 21: ...0V DC peak AC and 1000 Vrms mains isolated 200V DC peak AC and 200 Vrms CATII 500V DC peak AC and 500 Vrms mains isolated Absolute Maximum voltage each side to ground 300V DC peak AC and 200 Vrms CATI...

Страница 22: ...ies 170g 6 oz 160g 5 64 oz Including Accessories 1 01 kg 35 6 oz 1 83 kg 64 6 oz BNC Cable Length 1 2 m 47 inches 5 m 197 inches Length of Input Leads 39 cm 15 4 inches with extension leads 12 cm 4 7...

Страница 23: ...N2804A Performance Plots N2804 5A Differential Probes User s Guide 23 N2804A Performance Plots Figure 8 N2804A Derating Curve...

Страница 24: ...24 N2804 5A Differential Probes User s Guide N2804A Performance Plots Figure 9 N2804A Frequency Response Figure 10 N2804A Frequency Response when Inputs Driven in Common Mode CMRR...

Страница 25: ...N2804A Performance Plots N2804 5A Differential Probes User s Guide 25 Figure 11 N2804A Input Impedance Between Inputs Actual Figure 12 N2804A Input Impedance Between Inputs Simulated...

Страница 26: ...26 N2804 5A Differential Probes User s Guide N2804A Performance Plots Figure 13 N2804A Normalized Differential Step Response 50W...

Страница 27: ...N2805A Performance Plots N2804 5A Differential Probes User s Guide 27 N2805A Performance Plots Figure 14 N2805A Derating Curve...

Страница 28: ...28 N2804 5A Differential Probes User s Guide N2805A Performance Plots Figure 15 N2805A Frequency Response Figure 16 N2805A Frequency Response when Inputs Driven in Common Mode CMRR...

Страница 29: ...N2805A Performance Plots N2804 5A Differential Probes User s Guide 29 Figure 17 N2805A Input Impedance Between Inputs Actual Figure 18 N2805A Input Impedance Between Inputs Simulated...

Страница 30: ...30 N2804 5A Differential Probes User s Guide N2805A Performance Plots Figure 19 N2805A Normalized Differential Step Response 50W...

Страница 31: ...Bandwidth 200 MHz 1 MW 50W selectable input Keysight DSOS204A Display probe output Signal Generator Precision DC voltage source Amplitude accuracy less than or equal to 0 25 1 MW 50W selectable load S...

Страница 32: ...e 9500B set CH1 to the following settings and enable the output Waveform square wave Load 1 MW Amplitude 0 1V pk pk Frequency 1 kHz 5 On the oscilloscope select the amplitude measurement and record th...

Страница 33: ...tial Gain Accuracy in Table 10 on page 34 Bandwidth 15 Disable the 9500B s output 16 Connect the 50W BNC feed through adapter to the N9500B s CH1 active head 17 Attach the BNC f to banana post adapter...

Страница 34: ...est results as Bandwidth in Table 10 on page 34 24 Disable the 9500B s output and disconnect the probe input Performance Test Record Serial Number Certification Date Certification Temperature Tested B...

Страница 35: ...gory II is for measurements performed on circuits directly connected to the low voltage installation For example household appliances portable tools and similar equipment Measurement Category III CAT...

Страница 36: ...to the malfunctioning equipment Name and address of owner Product model number for example N2804A Product Serial Number for example MYXXXXXXXX Figure 20 Location of Serial Number Label Description of...

Страница 37: ...cting Keysight Technologies For technical assistance contact your local Keysight Call Center In the Americas call 1 800 829 4444 In other regions visit http www keysight com find assist Before returni...

Страница 38: ...38 N2804 5A Differential Probes User s Guide Returning the Probe for Service...

Страница 39: ...on oscilloscope 8 input resistance 21 inspecting 6 K Keysight Technologies contacting 37 L labels 7 M measurement category 10 12 20 mechanical characteristics 22 O Offset Adjust 15 oscilloscope compat...

Страница 40: ...40 N2804 5A Differential Probes User s Guide Index...

Отзывы: