N2804/5A Differential Probes User’s Guide
Probe Offset Adjustment
6
Set the oscilloscope to
Averaging
mode (x8 or higher) to reduce oscilloscope
noise.
7
Set the vertical scale of the oscilloscope to 100 mV/div (for N2804A) and
50 mV/div (for N2805A).
8
Using the offset adjustment tool that comes with the probe, adjust the probe
offset voltage to zero volts.