Performance Verification
4
N2795/6A User’s Guide
43
15
Perform Calibration for the
PORT 2
side of the Reference plane.
a
Press the
reflection
screen key.
b
Connect the open end of Calibration Standard to the
available end of the
PORT 2
SMA cable.
c
Select the
open
screen key under the
Reverse
group.
d
Wait until the VNA beeps indicating that it has completed the
task.
e
Connect short end of Calibration Standard to the available
end of the
PORT 2
SMA cable.
f
Select
short
screen key the
Reverse
group.
g
Wait until the VNA beeps indicating that it has completed the
task.
h
Connect load end of Calibration Standard to the available
end of the
PORT 2
SMA cable.
i
Select the
loads
screen key the
Reverse
group.
j
Press
broadband
screen key selection.
k
Wait until the VNA beeps indicating that it has completed the
task.
l
Press the
done loads
screen key.
m
You have just calibrated the other side of the reference plane.
16
Press
standards done
key.
17
Connect port two SMA cable to the non-pincher side of PV/DS
test board.
Содержание N2795/6A
Страница 1: ...Keysight N2795 6A Single Ended Probes User s Guide...
Страница 4: ...4 N2795 6A User s Guide Contents...
Страница 29: ...Characteristics and Specifications 2 N2795 6A User s Guide 29 Dimensions...
Страница 30: ...30 N2795 6A User s Guide 2 Characteristics and Specifications...
Страница 34: ...34 N2795 6A User s Guide 3 Performance Data Plots...
Страница 48: ...48 N2795 6A User s Guide 4 Performance Verification...
Страница 50: ...Index 50 N2795 6A User s Guide...