16
Keysight M9601A Startup Guide, Edition 2
Step 5: Verify Operation of the Module
To Update the Firmware
If a firmware update is available, perform the following procedure to update the
firmware of the M9601A modules.
1. Click Start menu > Keysight M960x Source Measure Unit > M960x SFP. The
“Connect to Instrument” window will open.
2. In the list on the window, highlight the M9601A modules to connect and click
the “Connect” button to launch the Keysight M960x PXIe Source Measure
Unit Soft Front Panel (the M960x SFP).
3. Confirm that the Access indicator on the M9601A front panel turns green.
Refer to
and
4. Select “Firmware Update...” from the Utilities menu. The “Firmware Update”
window will open.
5. Select the latest version of the firmware, if it is available.
If the select menu displays “No change”, there are no firmware updates.
6. Click the “Update” button at the bottom of the “Firmware Update” window to
start the firmware update.
When the firmware is updated successfully, the message appears in the
“Firmware Update” window.
7. Close the M960x SFP.
8. Restart your system; Power down the controller, then the chassis, and power
up the chassis, then the controller.