Step 3: Prepare and Power up the Signal Generator

Using Windows Tools

Only the current User or an Administrator can remotely log into
the instrument. To see who the current user of the instrument is,
press

Ctrl

+

Esc

on the instrument until you can view the current

user name on the Start menu. If no one is currently logged into
the instrument, any valid instrument user can remotely log in.

The instrument display appears on the screen of the remote computer. Since the
instrument front-panel keys are not available when using the instrument remotely,
three alternative methods are available to perform the functions of the front-panel
keys.

Windows Remote Desktop Options

Setting Remote Desktop Options

Step

Notes

1. On the Remote Desktop Connection menu, click
Options.

The Options dialog has several tabs.
Generally, the default settings are
correct.

2. Under the General tab, ensure that the computer
name, User name and Domain name are set correctly.

You may choose to enter the
password and save it for future
sessions, by selecting the Save my
password box.

M9384B Startup Guide

47

Содержание M9384B

Страница 1: ...Startup Guide Keysight M9384B VXG Microwave Signal Generator...

Страница 2: ...Pursuant to FAR 12 212 and 27 405 3 and Department of Defense FAR Supplement DFARS 227 7202 the U S government acquires commercial computer software under the same terms by which the software is cust...

Страница 3: ...ansient over voltages often associated with local AC mains connections Assume all measurement control and data I O connections are for connection to Category I sources unless otherwise marked or descr...

Страница 4: ...compliance with the Canadian Interference Causing Equipment Standard ICES 001 It also identifies the product is an Industrial Scientific and Medical Group 1 Class A product CISPR 11 Clause 4 South Ko...

Страница 5: ...ess protection This symbol indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use Forty years is the expected useful life...

Страница 6: ...This page has been intentionally left blank 6...

Страница 7: ...pplying Power 19 Servicing 19 Operating Conditions 19 For the AC DC Adapter 20 Unpack and Inspect the Instrument 22 Return an Instrument for Service 22 Verify M9384B VXG Shipment Contents 23 Step 2 Do...

Страница 8: ...and Use 38 Disk Drive Recovery Process 38 Updating the Software required after a recovery 41 Using Windows Tools 42 Navigating Windows Without a Mouse 42 Remote Desktop Using the Signal Generator Remo...

Страница 9: ...rability and troubleshoot your product Also includes an annotated block diagram PDF Embedded Help Provides user interface descriptions programming information tutorials SCPI commands and more HTML5 Da...

Страница 10: ...Introduction Documentation Access Documentation Access 10 M9384B Startup Guide...

Страница 11: ...ing the instrument you will need the following items A Keysight signal analyzer with the appropriate frequency range to measure the output of the M9384B VXG An Open End Wrench PN 8710 1761 to hold and...

Страница 12: ...This page has been intentionally left blank 12 M9384B Startup Guide...

Страница 13: ...Hz frequency reference Trig 1 SMB male Provides an input or output signal for trigger and events Trig 2 SMB male Provides an input or output signal for trigger and events Settled SMB male Outputs DDS...

Страница 14: ...MA female Deal with the quadrature phase component of I Q modulation required for external differential I Q I I Out SMA female Deal with the in phase component of I Q modulation provided by the intern...

Страница 15: ...Hz frequency reference EXT 1 SMA female Provides an input or output signal used for trigger and events SYNC OUT BNC female Provides an input or output signal used for trigger and events EXT CLK IN SMA...

Страница 16: ...or Current A 100 120 VAC 50 60 Hz 1200W Max 220 240 VAC 1300W Max Environmental Conditions Operating The table below lists the environmental conditions for the M9384B VXG For indoor use only Altitude...

Страница 17: ...ce Verify M9384B VXG Shipment Contents ESD Electrostatic discharge ESD can damage or destroy electronic components Use a static safe work station to perform all work on electronic assemblies The figur...

Страница 18: ...to avoid having the heated discharge of one instrument now becoming the cooling intake air for another instrument Another area of concern is verification that the maximum ambient operating temperatur...

Страница 19: ...II and IV The input of this instrument cannot be connected to the mains The Mains wiring and connectors shall be compatible with the connector used in the premise electrical system Failure to ensure a...

Страница 20: ...It disconnects the mains circuits from the mains supply before other parts of the instrument The front panel switch is only a standby switch and is not a LINE switch Alternatively an externally insta...

Страница 21: ...ed 10 percent of the nominal supply voltage The AC DC adapter is for indoor use only Never use a modified or damaged charger Use the original AC DC adapter ONLY The AC DC adapter is designed for use i...

Страница 22: ...ecautions are taken Store the instrument in an anti static envelope when not in use Visit www keysight com find tips for information on preventing damage to your Keysight equipment Return an Instrumen...

Страница 23: ...hem in a separate protective bag 5 On the shipping label write ATTENTION REPAIR DEPARTMENT and the RMA number 6 In your correspondence refer to the instrument by model number Verify M9384B VXG Shipmen...

Страница 24: ...This page has been intentionally left blank 24 M9384B Startup Guide...

Страница 25: ...nstrument software upgrades application information and product information visit www keysight com find M9384B For the worldwide contact information for repair and service visit www keysight com find...

Страница 26: ...This page has been intentionally left blank 26 M9384B Startup Guide...

Страница 27: ...ctices The following are the recommended best practices to ensure proper and safe operating conditions Ensure that adequate clearance is provided around all instrument vents both air intake vents and...

Страница 28: ...er sides lower front and bottom of the instrument Prepare and Power up the Instrument 1 Make sure that the line cord is plugged into a grounded outlet to establish earth ground 2 Make sure that the in...

Страница 29: ...To search for software updates for your product visit www keysight com find M9384B Application Keyboard Shortcuts M9384B VXG Microwave Signal Generator allows you to move the application window by pr...

Страница 30: ...ument Add or change user accounts and passwords Change Firewall settings Change Windows settings for example using Device Manager Change the time and date Run any application Instrument Accounts The d...

Страница 31: ...Windows Systems settings you may have to reinstall the Windows system and instrument applications using the Instrument Recovery process Settings that can be Changed You may change the following Windo...

Страница 32: ...ave the remaining selections unchanged Settings that must not be Changed Avoid changing the settings described below available from the Windows Control Panel Changes to these settings may degrade inst...

Страница 33: ...braries The NET Framework or any Hotfixes or Service Packs for the NET Framework The Microsoft Visual J NET Redistributable Package 1 1 Programs that begin with Keysight The Adobe Acrobat Reader Modif...

Страница 34: ...DHCP You can change the IP Address and Gateway as you desire You must be logged in as an administrator default password Keysight4u to make changes Connecting to a Network Shared Folder The instrument...

Страница 35: ...e anti virus software Anti Virus Software and Firewalls No third party anti virus software is shipped with the M9384B VXG It is recommended that you install an anti virus software if your M9384B VXG i...

Страница 36: ...then you should manually update Windows periodically Be aware that downloading and installing Windows Updates can be network and CPU usage intensive impacting the instrument performance and some Windo...

Страница 37: ...ettings as provided by Microsoft However System Restore is not always 100 successful Therefore it is not recommended that you rely on System Restore to protect your instrument System Restore has not b...

Страница 38: ...the C drive is the only part of the instrument software that is restored by the Instrument Recovery process You must reload any other software that you have added in the instrument The D partition is...

Страница 39: ...Instrument Recovery overwrites the contents of the C partition It does not affect the D or E partitions It is recommended that you use a regular back up strategy Your IT department may already have a...

Страница 40: ...ow the on screen instructions to recover the image of the C drive Press 2 then press Enter to select the recovery 4 A warning message appears Press Enter to start the recov ery which may take up to 25...

Страница 41: ...are available at www keysight com find M9384B Installing the Software Step 1 Log out of the default user instrument Click Start Select Change account settings Sign out 2 At the log in prompt enter Us...

Страница 42: ...n this section are Microsoft Windows 10 features The discussion provided here gives some guidelines for using the capabilities with the instrument Refer to the Windows 10 help documentation for more i...

Страница 43: ...Alt Esc Cycles through items in the order that they had been opened Backspace In My Computer or Windows Explorer move up one level In Internet Explorer works like the BACK arrow key Ctrl Left arrow Mo...

Страница 44: ...ty and also how to set up a computer running any version of Microsoft Windows as a Remote Desktop Host Setting Up a Remote Desktop Operation Basic Setup for Remote Desktop Operation To perform this op...

Страница 45: ...software The procedure described here may be changed by Microsoft at some future time Installing the Client Software Step Notes 1 When the Welcome Screen appears click Perform additional tasks 2 From...

Страница 46: ...of the instrument After setting up the remote computer for Remote Desktop Connectivity as described in Setting Up a Remote Desktop Operation you are ready to start a Remote Desktop session Starting a...

Страница 47: ...he remote computer Since the instrument front panel keys are not available when using the instrument remotely three alternative methods are available to perform the functions of the front panel keys W...

Страница 48: ...accessible Under Colors set this to 16 bits If you oper ate Remote Desktop with greater color depth e g 32 bit your windows may have a dif ferent appearance than they do on the instru ment display bec...

Страница 49: ...ick Start Disconnect Click Disconnect You are prompted to confirm that you want to disconnect Windows Shortcuts and Miscellaneous Tasks This section provides a list of windows shortcuts key combinatio...

Страница 50: ...og box items Shift Tab In a dialog open a list box Alt Down Arrow In a dialog list box or check box select or deselect one item at a time Shift Up Arrow Shift Down Arrow In My Computer expand a select...

Страница 51: ...e auto hide option is selected then press Select to toggle the check box state Click OK This applies the change and closes the dialog box Trigger Routing The M9384B VXG supports the following triggeri...

Страница 52: ...This page has been intentionally left blank 52 M9384B Startup Guide...

Страница 53: ...t make sure that the instrument is powered on and all cabling is correct 1 Launch the M9384B application 2 Click the Gear icon on the top right corner to open the System dialog window 3 Select Diagnos...

Страница 54: ...This page has been intentionally left blank 54 M9384B Startup Guide...

Страница 55: ...on the Keysight Signal Analyzer 2 Connect a high quality male 2 4 mm to SMA male cable between the Front Panel RF Out connector and the RF In connector on the Keysight Signal Ana lyzer 3 Torque the co...

Страница 56: ...of the display set RF Out All to On by selecting the switch 5 On the Keysight Signal Analyzer select AUTO TUNE in the Frequency menu You should see the following display on your Keysight Signal Analy...

Страница 57: ...Step 6 Installation is Complete Proceed to program your product by means of the SCPI programming interface M9384B VXG Microwave Signal Generator Startup Guide 57...

Страница 58: ...This page has been intentionally left blank 58 M9384B Startup Guide...

Страница 59: ...This information is subject to change without notice Keysight Technologies 2019 Edition 1 0 July 2019 M9384 90001 www keysight com...

Отзывы: