31
Keysight M9217A Startup Guide
Figure 9
Configuring the digitizer
4
Set the calibrator to output -0.235 V.
5
On the M9217A PXIe 2-CH Digitizer SFP, click on the
Acquire Data
tab. Click on
Single
to
measure. The result will be displayed as shown in
.
Table 3
Configuration for digitizer
Settings
Description
Specification
Channels
Channel
1
Range
500 mV (±250 mV)
Coupling
DC
Acquisition
Sample rate
20 MSa/s
Number of records
1
Record size
20000
Trigger holdoff
0 sec
Pre-trigger samples
0
Trigger delay
0
Trigger source
Immediate