Safety Information
Installation and Configuration
Providing Adequate Clearance
Apertures in the top, bottom, front, rear, and along the right side of the
chassis facilitate power supply and module cooling. Air enters through a fan
intake in the bottom of the chassis for module cooling. It then exits through
the upper sections at the right side, back, and through the top, as shown in
. Air cooling the power supply enters the front of the chassis, which is
shown in
on page 4, then exits through the rear of the chassis, which
is shown in
on page 5. Place the chassis on a bench top or in an
instrument rack so that the fans (air intakes) and the air outlet apertures along
the right side, the top, and the back of the chassis have adequate ventilation.
Provide at least 44.5 mm (1.75 in.) clearance above, behind, and on the sides
of the unit for adequate venting. High-power applications may require
additional clearance.
Figure 8
M9005A Module Cooling Airflow Side View
Failure to provide adequate clearances may result in thermal
related failures in the chassis or modules.
1 Air Outlets
2 Air Intake
2
1
Do not block air intake. This could lead to overheating.
Содержание M9005A
Страница 1: ...Keysight M9005A PXIe Chassis User Guide...
Страница 2: ......
Страница 6: ...vi...
Страница 9: ...Keysight M9005A PXIe Chassis User Guide ix 6 Pinouts Hybrid Slot Pinouts 51 Index...
Страница 10: ...x Keysight M9005A PXIe Chassis User Guide...
Страница 22: ...12 Keysight M9005A PXIe Chassis User Guide Getting Started M9005A Chassis Backplane Overview...
Страница 42: ...32 Keysight M9005A PXIe Chassis User Guide Installation and Configuration Getting Started With The System...
Страница 48: ...38 Keysight M9005A PXIe Chassis User Guide Troubleshooting...
Страница 64: ...54 Keysight M9005A PXIe Chassis User Guide Pinouts...
Страница 67: ...57...