2.2.4: Connecting a remote controller PC to the chassis
A remote controller is an external, Windows-based PC that connects to
the chassis through a PCIe cable. The remote controller can be a desktop
PC, a laptop PC, or a rack-mounted PC. If your configuration contains a
Keysight Cable Interface module, follow the procedure below.
1 Locate slot 1 in the chassis, with the
icon above the slot.
2 Set the chassis controller slot power supply switch to the right-hand
position. This provides power to slot 1 for the benefit of the PCIe
interface card.
3 Remove the interface module from its protect bag. See “
4 On the module, set both S301 switches to the Host (right-hand)
position and set the S201 rocker switch to the left-hand position.
5 Install the Cable Interface module into the chassis:
•
While holding the module by the injector/ejector handle and making
sure the injector/ejector handle is pushed down in the unlatched
(downward) position, slide the module into chassis, using the slot
guides (top and bottom).
•
Sliding the module into position, when you begin to feel resistance,
pull up on the injector/ejector handle to fully inject the module into
the chassis backplane connectors.
•
Tighten the module retaining screws (top and bottom) and torque
them to 5 Lb-In (0.57 N-m).
NOTE
The following procedure addresses using a cabled PCIe interface
between the chassis and an external host computer. However, if you
intend to use a system module to control a subordinate downstream
chassis or RAID configuration:
1. Install the module in an x8 hybrid slot in the PXIe chassis.
2. Reverse the switch settings from those noted in this procedure: On the
module, set both the S301 switches to “Host” and set the S201 rocker
switch to the left-hand position. On the chassis backplane, set the
controller slot power-supply switch to the left.
Содержание M5201A
Страница 1: ...0 3 H RZQFRQYHUWHU 0RGXOHV RXU 5 67 5783 8...
Страница 8: ...8 M5201A PXIe Downconverter Modules Startup Guide...
Страница 12: ...12 M5201A PXIe Downconverter Modules Startup Guide Contents...
Страница 26: ...26 M5201A PXIe Downconverter Modules Startup Guide 1 Overview on M5201A PXIe Downconverter modules...
Страница 66: ...66 M5201A PXIe Downconverter Modules Startup Guide 3 Setting up the M5201x software...
Страница 72: ...72 M5201A PXIe Downconverter Modules Startup Guide 4 Troubleshooting and Safety information...
Страница 73: ...M5201A PXIe Downconverter Modules Startup Guide 73...