Testing Performance
2
DSOX1204A/G Oscilloscopes Service Guide
21
d
Press the
[Acquire]
key.
e
Then press the
Acq Mode
softkey and select
Averaging
.
f
Then press the
#Avgs
softkey and set it to 64.
Wait a few seconds for the measurement to settle.
4
Add a measurement for the average voltage:
a
Press the
[Meas]
key.
b
Press
Source
; then, turn the Entry knob (labeled
on the front panel) to
select the channel you are testing.
c
Press
Type:
; then, turn the Entry knob to select
Average - Full Screen
, and press
Add Measurement
.
5
Read the “current” average voltage value as V1.
6
Use the BNC tee and cables to connect the oscilloscope calibrator/power
supply to both the oscilloscope and the multimeter (see