22
DSOX1204A/G Oscilloscopes Service Guide
2
Testing Performance
Figure 1
Setting up Equipment for DC Vertical Gain Accuracy Test
7
Adjust the output so that the multimeter reading displays, as close as possible,
the first Volts/div calibrator setting value in
8
Wait until the measurement settles.
9
Read the “current” average voltage value again as V2.
10
Calculate the difference V2 - V1.
Oscilloscope
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
BNC (f) to dual
bananna adapter
Содержание InfiniiVision DSOX1204A