6 List Sequence Measurements
Source
default directory D: VARB. The SCPI command supports using either “NVWFM” MSUS or specifying a full
path. For more information, see
Memory Subsystem (Remote Command Only)
If a file of the same name already exists within ARB memory, it is overwritten. If you wish to load two
segments of the same name, you must rename one of the segments before loading it into ARB memory. To
rename a segment, you can either use Windows File Explorer, or the :MMEMory:COPY command.
Key Path
Source, Modulation Setup, ARB, Select Waveform, Segments on Hard Disk
Remote Command
:SOURce:RADio:ARB:LOAD <string>
Example
:SOUR:RAD:ARB:LOAD “D: VARB\testwaveform.bin”
or
:SOUR:RAD:ARB:LOAD “NVWFM:testwaveform.bin”
Notes
Because loading the file involves a delay of unpredictable length, this command should be followed by
the query *OPC?, which holds off subsequent commands until the loading operating is complete.
<string> – specifies the path name of the file to load from the HDD into ARB memory. It could be a
<full path + filename>, or <“NVWFM” MSUS + colon + filename>.
When in Sequence Analyzer mode, and Include Source is Yes, an attempt to load a file to ARB
memory will be rejected with an error.
When Include Source is Noand if there is insufficient free ARB memory to load the selected
waveform, an error is generated. .
If you specify a file over SCPI, but the file is not at the specified location, an error is generated.
If you try to load a waveform file but the file contains less than 500 IQ sampes, an error is generated.
If you try to load a Signal Studio waveform “*.wfm” which contains invalid waveform header, an error
is generated.
If the ARB is ON when you load afile to ARB memory or delete a file from ARB memory, the playing
waveform segment may not keep phase continuity during the ARB memory operation. The waveform
will be replayed after the ARB operation is finished.
ARB can be loaded into ARB memory even required licenses do not present on the instrument. In this
case, a GUI only warning message –800, “Operation complete; Loaded <filename> successfully, but
no license <required licenses> installed”. User can install required licenses according to <required
licenses> string to license it, or multi-pack license it.
Initial S/W Revision
A.05.00
Load All To ARB Memory
Allows you to load all the segment files within the currently selected directory into ARB memory. If a file of
the same name already exists within ARB memory, it is overwritten. If you wish to load two segments of the
same name, you must rename one of the segments before loading it into ARB memory. To rename a
segment, you can either use Windows File Explorer, or the :MEMory:COPY command.
Key Path
Source, Modulation Setup, ARB, Select Waveform, Segments on Hard Disk
Remote Command
:SOURce:RADio:ARB:LOAD:ALL <string>
Example
:SOUR:RAD:ARB:LOAD:ALL “D: varb”
Notes
<string> - specifies the directory on the HDD to load the files into ARB memory from.
When in Sequence Analyzer mode, and Include Source is Yes, an attempt to load all files from a
Sequence Analyzer User's & Programmer's Reference
623
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...