Keysight E5063A Network Analyzer 25
Troubleshooting
Booting Process Troubleshooting
Step 4.
System Initialization Screen.
The system initialization screen is displayed. If the display whites out, is
entirely blue or a dialog box appears, a mass storage problem is suspected.
Try to perform the mass storage recovery procedure.
If the E5063A was turned off without proper shutdown process,
Microsoft
Scandisk
runs while the windows boot screens are displayed. If a serious
problem is found during the scan, execute the system recovery. For details
on executing the system recovery, refer to the
Installation Guide
. If the
operating system still does not boot up properly after reinstallation,
replace the HDD.
The operating system automatically checks the device drivers which are
necessary for use in the E5063A and are installed in the system before the
E5063A is shipped from Keysight factory. If the operating system does not
detect them, a message box is displayed. In this case, install the device
driver.
If the message “Will Shut Down in Five Seconds” is displayed and
shutdown occurs, the A51 DSP board has failed to start up. The following
messages may be displayed before the shutdown occurs:
“Fatal Error: Failed to Initialize DSP Driver”
or “Fatal Error: Failed to Initialize DSP”
This message indicates that the DSP board does not work or is not
properly connected to the DSP DSP board.
“Fatal Error: Failed to Update DSP Code”
If this happens, the DSP board has failed to write the DSP program into the
flash ROM when the firmware was first installed or after being updated to
the latest version. A problem with the A51 DSP board or A60 CPU Module
is suspected.