Keysight E5063A Network Analyzer 31
Troubleshooting
Function Specific Troubleshooting
To Check the Front Panel
Procedure
Step 1.
Press
key.
Step 2.
Click
Service Menu
-
Test Menu
-
Front Panel
. This opens the Front Panel
Test window as shown in
.
Figure 1-15
Front Panel Test window 1/2
Step 3.
Press the front panel keys. Red ticks will appear on the Front Panel Test
window whenever the corresponding key on the front panel is pressed as
shown in
. Turn the rotary knob clockwise or counterclockwise.
Red ticks will appear on the left of the rotary knob on the Front Panel Test
window when you turn counterclockwise and on the right of the rotary knob
when you turn clockwise.
Handler
Interface
The Handler Interface port can be used to
transfer comparator decision output data
and perform timing synchronization with
an external handler.
Refer to
Table 1-1
Major functions and troubleshooting procedures
Function
Description
Troubleshooting