Calibration Procedures
1
Keysight E3632A Service Guide
21
Automating Calibration Procedures
You can automate the complete verification procedures outlined in this chapter if
you have access to programmable test equipment. You can program the
instrument configurations specified for each test over the remote interface. You
can then enter readback verification data into a test program and compare the
results to the appropriate test limit values.
You can also enter calibration constants from the remote interface. Remote
operation is similar to the local front-panel procedure. You can use a computer to
perform the adjustment by first selecting the required setup. The calibration value
is sent to the power supply and then the calibration is initiated over the remote
interface. The power supply must be unsecured prior to initiating the calibration
procedure. A Keysight BASIC program for calibration over the GPIB interface is
listed at the end of this chapter.
For further details on programming the power supply, see chapters 3 and 4 in the
Keysight
E3632A User's Guide.
Содержание E3632A
Страница 1: ...Keysight E3632A DC Power Supply Service Guide ...
Страница 12: ...Keysight E3631A Service Guide 14 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...16 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 68: ...2 Service 70 Keysight E3632A Service Guide ...
Страница 70: ...72 Keysight E3632A Service Guide Component locator diagram for the front panel top ...
Страница 71: ...Keysight E3632A Service Guide 73 Component locator diagram for the main board assembly top serial MY53xx6xxx ...
Страница 72: ...74 Keysight E3632A Service Guide Component locator diagram for the front panel top serial MY53xx6xxx ...