Calibration Procedures
1
Keysight E3632A Service Guide
23
Test Considerations
To ensure proper instrument operation, verify that you have selected the correct
power-line voltage prior to attempting any test procedure in this chapter. Refer to
the E3632A User’s Guide for more information.
For optimum performance verification, all test procedures should comply with the
following recommendations.
– Assure that the calibration ambient temperature is stable and between 20 °C
and 30 °C.
– Assure ambient relative humidity is less than 80%.
– Allow a 1-hour warm-up period before verification or calibration.
– Keep cables as short as possible, consistent with the impedance requirements.
CAUTION
The tests should be performed by qualified personnel. During performance
verification tests, hazardous voltages may be present at the outputs of the
power supply.
Содержание E3632A
Страница 1: ...Keysight E3632A DC Power Supply Service Guide ...
Страница 12: ...Keysight E3631A Service Guide 14 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...16 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 68: ...2 Service 70 Keysight E3632A Service Guide ...
Страница 70: ...72 Keysight E3632A Service Guide Component locator diagram for the front panel top ...
Страница 71: ...Keysight E3632A Service Guide 73 Component locator diagram for the main board assembly top serial MY53xx6xxx ...
Страница 72: ...74 Keysight E3632A Service Guide Component locator diagram for the front panel top serial MY53xx6xxx ...