Service Information
5
Keysight E3630A Operating and Service Manual
51
Rated output and ammeter accuracy
6
Connect two 40
Ω
20 W load resistors across both of the 20 V outputs of the
supply and set the ±20 V VOLTAGE control for ±20 V outputs. (All supplies
must be fully loaded while checking the rated output voltage and current of
each supply.)
7
Connect the test setup shown
to the +6 V output. Make the total
resistance of R
L
and the current sampling resistor equal to 2.4
Ω
to permit
operating the output at full load. R
L
should have a power rating of at least
20 watts.
8
Close the switch and set the +6 V VOLTAGE control so that the DVM indicates a
voltage drop across the current sampling resistor that corresponds to a current
of 2.5 amps.
9
Push the +6 V METER switch and verify that the front panel ammeter indication
is within ±(0.5% + 2 counts) of 2.5 amps (2.47 A to 2.53 A).
10
Check the rated output and ammeter accuracy of the +20 V and –20 V
supplies similarly by connecting the test setup of
to each output in
turn. For each 20 V supply: make the total resistance of R
L
and the current
sampling resistor 40
Ω
, set the ±20 V VOLTAGE control for a current indication
on the DVM of 0.5 A, check that the panel meter indication is within ±(0.5%+2
counts) of 0.5 A (0.48 A to 0.52 A).
11
Disconnect all loads from the supply.
12
Connect the test setup shown in
to the +20-volt output. Substitute
a short for R
L
and leave the load circuit switch open.
13
Set the voltage of the ±20 V supplies to 20 volts.
14
Close the load switch and determine the current flow through the current
sampling resistor (meter shunt) by measuring its voltage drop with the DVM.
The current should be 0.55 A ±5% (0.5225 A to 0.5775 A).
15
Check the current limit of the –20 V supply in the same way. Its short-circuit
current should also be 0.55 A ±5% (0.5225 A to 0.5775 A).
16
Connect the test setup shown in
to the +6 V output. Close the
switch, set the total resistance of R
L
and the current sampling resistor to an
initial value of 2.4
Ω
or greater, and set the output voltage to 6 volts.
Содержание E3630A
Страница 1: ...Triple Output Power Supply Keysight Model E3630A Operating and Service Manual ...
Страница 10: ...10 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 64: ...6 Component Locator Diagrams 64 Keysight E3630A Operating and Service Manual Main Board Top ...
Страница 65: ...Component Locator Diagrams 6 Keysight E3630A Operating and Service Manual 65 Main Board Bottom ...
Страница 66: ...6 Component Locator Diagrams 66 Keysight E3630A Operating and Service Manual Display Controller Top ...
Страница 67: ...Component Locator Diagrams 6 Keysight E3630A Operating and Service Manual 67 Display Controller Bottom ...
Страница 68: ...6 Component Locator Diagrams 68 Keysight E3630A Operating and Service Manual 7 Segment Top ...
Страница 69: ...Component Locator Diagrams 6 Keysight E3630A Operating and Service Manual 69 7 Segment Bottom ...