2 Installing the N1000A
3. Remove all external devices and signals from the module. The module cal-
ibration uses known signal levels in the instrument. Introducing an outside
signal source interferes with the calibration factors and decreases cal-
ibration accuracy.
4. In the dialog box, select a tab that represents the module that you want to
calibrate.
5. Calibrations with a green check mark are valid. Select any items without a
green check mark or that show any other icon and click
Calibrate
to starts
the module calibration.
6. Follow any displayed instructions. After the calibration completes, the dia-
log will show the status of the calibration.
34
N1000A Quick Start Guide