Keysight B2980B User’s Guide, Edition 1
107
How to Perform Measurement
Resistance Measurement
Figure 3-15
Grounded Device Measurement using N1414A
For the grounded device measurement, set the N1414A control switch to the
PUSH position. Also set the voltage source Low terminal state to FLOATING.
The connection shown in
applies the voltage to the Common terminal
from the Voltage Source. The voltage up to
500 V can be applied to the Common
terminal.
Procedure
You can perform the resistance measurement as follows.
Step 1. Set the resistance calculation mode, Vs/Im or Vm/Im. See
Step 2. Press the
View
key and select the Meter View function key to display the Meter
view.
Step 3. Press the
OHMS (R)
assist key to set the resistance measurement mode.
Step 4. Set the current measurement range and the output voltage. See
Input
Triaxial cable
Triaxial bulkhead connector
N1414A adapter
High voltage test lead
Low
Common is connected to Voltage Source High
through the N1414A adapter.
Guard shield
Control switch:
PUSH
position
Shield
R
X
Input
R
X
Low
Guard
shield
Shield
A
V
Common
V
S
V
M
= 0.01*V
S
V
M
Voltage Source Low Terminal State:
FLOATING
100:1
Содержание B2980B Series
Страница 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Страница 16: ......
Страница 24: ......
Страница 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Страница 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Страница 130: ...130 Keysight B2980B User s Guide Edition 1 How to Perform Measurement Measurement Considerations...
Страница 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Страница 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...