Overview
Description
The Keysight 537A coaxial frequency meter is used for measuring frequency in the 0.96 to
12.4 GHz range. Accuracy is obtained from the spiral scale. The meter contains a coaxial line
coupled to a quarter- wavelength coaxial resonant cavity loaded with capacitance at the bottom of
the band. At resonance, power is absorbed by the cavity, producing a dip in the coaxial line power.
The frequency at which this dip occurs is read directly from the dial in gigahertz.
Tuning is accomplished by moving a piston. A precision lead screw, which is spring-loaded to
prevent backlash, is used to position the piston in the cavity. Due to the design of the cavity, its
electrical length is extended at lower frequencies, and there are no spurious modes.
Specifications
The instrument specifications are provided in
Table 1
Specifications
Characteristic
Value
Frequency range
3.7 to 12.4 GHz
Reflection coefficient off-resonance
0.33 (2.0 SWR, 9.5 dB return lost)
Calibration increments
10 MHz
Dimensions
Height
Diameter
[a]
[a] Width (including connector): 4-5/8in (118 mm)
5-3/4 in (146 mm)
3-1/2 in (89 mm)
Weight
3-1/2 lb (1.6 kg)
Dip at resonance
1 dB (minimum)
Dial accuracy
[b]
[b] Overall accuracy: ±0.17%, which includes allowance of ±0.02% for 0 to 100% relative humidity, ±0.0016 per ×C from 13 to 33×C and 0.03%
backlash.
±0.1%
Содержание 537A
Страница 1: ...Keysight 537A Coaxial Frequency Meter Operating Note...
Страница 6: ...6 Keysight 537A Operating Note THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 8: ...8 Keysight 537A Operating Note THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 10: ...10 Keysight 537A Operating Note THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 12: ...12 Keysight 537A Operating Note THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 22: ...Calibration Procedure 22 Keysight 537A Operating Note THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...