2
Operating Your Instrument
70
Keysight 53147A/148A/149A Operating and Programming Guide
Summary of the Measurement Sequence
1
Connect the power cord to the instrument and to an appropriate power source
(or, if the instrument has the Battery option, turn on the Battery Power switch),
and then press and release the
POWER
button on the front panel.
2
Connect the input signal to the appropriate input connector (Counter Channel
1, Counter Channel 2, Power Meter input, or DVM input).
3
Connect an external reference signal to the External Reference connector on
the back panel (if desired).
4
Press the
Chan Select
key, the
Display Power
key, and/or the
Display DVM
key to select the measurement function (if necessary).
5
Press the
Freq Offset
key and/or the Power Meter
Offset On/Off
key(s) to
enable offset measurements (if desired), and then use the
Freq Offset
(
Shift
+
Resol
) and/or
Pwr Offset
(
Shift
+
Avg
) keys to enter the offset values.
6
Use the
Resol
,
Rate
(
Shift
+
+/–
), and
Avg
keys to configure the display.
7
Use the
Menu
(
Shift
+
Reset/Local
) key to set the reference-oscillator source,
to select the Channel 1 low-pass filter, to configure the Counter’s response to
frequency modulation, to select a Power Meter head, and/or to select a
previously saved set of user settings.
8
If you intend to operate the instrument remotely using the GPIB,
use the
Menu
(
Shift
+
Reset/Local
) key to set the GPIB address.
9
If you intend to operate the instrument remotely using the serial interface, use
the
Menu
(
Shift
+
Reset/Local
) key to adjust the serial port Baud rate.
NOTE
The internal Reference Oscillator receives power whenever the power cord is
connected to a power source. Therefore, if the instrument has just been
connected to a power source, the frequency of the reference signal may drift
until the oscillator stabilizes. Specifications for the stability of the standard
internal timebase and the optional Oven Timebase are provided in Chapter 3.