3
Programming
258
Keysight 53147A/148A/149A Operating and Programming Guide
*OPC? (Operation Complete Query)
COMMAND
*OPC?
This query causes the instrument to place a response in the output queue when a
triggered action is complete. This allows synchronization between a controller and
the instrument using the
MAV
bit in the Status Byte Register. (Note that this query
does not actually “read” a state, as most queries do.)
See the section titled
that uses this command.
QUERY
RESPONSE
Single ASCII-encoded byte, 1.
RELATED
FRONT-PANEL
KEYS
None
*PRE (Parallel Poll Enable Register)
*PRE? (Parallel Poll Enable Register Query)
COMMAND
*PRE<NRf>
*PRE?
Sets or queries the value of the parallel poll enable register.
QUERY
RESPONSE
<NRf>
RANGE
0-255
COMMENTS
– The parallel poll enable register is eight bits wide and has the same bit
definitions as the status byte.
– The status byte and parallel poll enable registers are anded together; the
result determines the value of
*IST
(
TRUE
or
FALSE
).
NOTE
The *OPC? query does not in any way affect the OPC bit in the Standard Event
Status Register.