CONFigure[:VOLTage]:{AC|DC} [{<
range
>|AUTO|MIN|MAX|DEF} [, {<
resolution
>|MIN|MAX|DEF}]]
Sets all measurement parameters and trigger parameters to their
for AC or DC voltage meas-
urements. Also specifies the range and resolution.
The maximum range parameter (MAX) is 1000 V. However, the SAFETY LIMIT on the front
and rear HI/LO input terminals is 750 VAC (rms). The rms voltage is waveform dependent.
A sine wave is limited to 750 VAC (rms), but a 1000 Vpk square wave is safe. Connection
to AC mains power is further limited to CAT II (300 V). See
for more information on the safety features and the safe operation of this
instrument.
Parameter
Typical Return
<
range
>: {100 mV|1 V|10 V|100 V|1000 V}. Default: AUTO (autorange).
(none)
<
resolution
> (AC): optional and ignored; fixed at 6½ digits.
<
resolution
> (DC): See
or
The
default is equivalent to 10 PLC.
Specify <
resolution
> in the measurement's units (V, A, Hz, Ω, and so on).
Configure AC voltage measurements using the 100 V range. Make and read two measurements:
CONF:VOLT:AC 100
SAMP:COUN 2
READ?
Typical Response: +8.5401,+8.5401
Configure DC voltage measurements using the 10 V range with 1 mV resolution. Then make a meas-
urement using an external trigger with positive slope (rising edge) and read the measurement:
CONF:VOLT:DC 10,0.001
TRIG:SOUR EXT;SLOP POS
INIT
FETC?
Typical Response: +4.2700
l
You can let autoranging select the measurement range, or you can manually select a fixed range.
Autoranging conveniently selects the range for each measurement based on the input signal. For fast-
est measurements, use manual ranging (autoranging may require additional time for range selection).
l
Autoranging (AUTO or DEFault), will generate an error if you specify a <
resolution
> because the instru-
ment cannot accurately resolve the integration time (especially if the input continuously changes). If
your application requires autoranging, specify DEFault for the <
resolution
> or omit the <
resolution
>
altogether.
l
Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of
range.
l
If the input signal is greater than can be measured on the specified manual range, the instrument dis-
plays the word
Overload
on front panel and returns "9.9E37" from the remote interface.
SCPI Programming Reference
252
Keysight True
volt
Series Operating and Service Guide
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...