4-8 Keysight R, Q, U, V, and W11644A
Use, Maintenance, and Care of the Devices
Connections
4-
Figure 4-6
“X” Screw Pattern
Inspecting a Flange Connection
Inspect the flange connection as follows:
1.
Place an electric light or white paper behind the connection.
2.
Check the flange matings for any gap. A good connection has no gaps
between the connected waveguide flanges, and the waveguide walls are
flush. There is no step or offset.
3.
Ensure that all four screws are equally tight.
The most common cause of measurement error is a poor connection.