Keysight R, Q, U, V, and W11644A 2-1
Specifications
Environmental Requirements
2
Specifications
Environmental Requirements
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the
operating temperature is a critical factor in performance. During a
measurement calibration, the temperature of the calibration devices must be
stable and within the range shown in
Table 2-1
Environmental Requirements
Parameter
Limits
Temperature
Operating
a
a. The temperature range over which the calibration standards maintain performance to their specifications.
+20 °C to +26 °C
Storage
−
40 °C to +75 °C
Error-corrected range
b
b. The allowable network analyzer ambient temperature drift during measurement calibration and during mea-
surements when the network analyzer error correction is turned on. Also, the range over which the network
analyzer maintains its specified performance while correction is turned on.
±
1 °C of measurement calibration temperature
Relative humidity
Type tested, 0% to 95% at 40 °C, non-condensing
IMPORTANT! Avoid unnecessary handling of the devices during calibration
because your fingers act as a heat source and may increase the
temperature of the device.
Содержание 11644A Q
Страница 3: ...3...
Страница 4: ...4...
Страница 12: ...1 6 Keysight R Q U V and W11644A General Information When to Calibrate 1...
Страница 24: ...3 8 Keysight R Q U V and W11644A User Information Changing the Wavelength Shim Calibration Definition 3...
Страница 34: ...4 10 Keysight R Q U V and W11644A Use Maintenance and Care of the Devices Handling and Storage 4...
Страница 40: ...6 2 Keysight R Q U V and W11644A Troubleshooting Troubleshooting Process 6 Figure 6 1 Troubleshooting Flowchart...
Страница 54: ...7 12 Keysight R Q U V and W11644A Replaceable Parts Introduction 7...
Страница 61: ...Installation Note Xxxxx xxxxx 3...