Keysight R, Q, U, V, and W11644A 5-3
Performance Verification
Performance Test
Performance Test
Termination Return Loss Measurement
Use this test to check the performance of the terminations in this kit.
Figure 5-1
Return Loss Test Setup
1.
Connect the equipment as shown in
. Turn on and preset the
network analyzer. Let the equipment warm up for at least one hour.
2.
Set the appropriate start frequency to:
—
26.5 GHz (R-band)
—
33 GHz (Q-band)
—
40 GHz (U-band)
—
50 GHz (V-band)
—
75 GHz (W-band)
3.
Set the appropriate stop frequency to:
—
40 GHz (R-band)
—
50 GHz (Q-band)
The calibration kit definitions must already be loaded in the analyzer. Refer
to your analyzer user’s guide for information on how to load the calibration
kit definitions.