84
Keysight 1000 Series Oscilloscopes User’s Guide
3
Capturing Data
7
Press
Setting
and turn the entry knob to adjust the width setting.
To set up video triggers
Video triggering is used to trigger on fields or lines of NTSC, PAL, or SECAM
standard video waveforms.
When the video trigger mode is selected, the trigger coupling is set to AC.
1
Press
[Menu]
.
2
In the Trigger menu, press
Mode
.
3
Continue pressing the
Mode
softkey or turn the entry knob to select “Video”.
4
Then, either push the entry knob or press
Mode
again.
5
Press
Polarity
to toggle between:
6
Press
Sync
and continue pressing the softkey or turn the entry knob to select
what to trigger on:
Positive pulse less than the width setting.
Negative pulse greater than the width setting.
Negative pulse less than the width setting.
Normal polarity — trigger on the negative edge of the sync
pulse.
Inverted polarity — trigger on the positive edge of the sync
pulse.
NOTE
Normal Polarity Sync triggers always occur on negative-going horizontal sync pulses. If the
video waveform has positive-going horizontal sync pulses, use the Inverted Polarity selection.
All Lines
Trigger on all lines.
Содержание 1000 Series
Страница 12: ...12 Keysight 1000 Series Oscilloscopes User s Guide Contents ...
Страница 16: ...16 Keysight 1000 Series Oscilloscopes User s Guide Tables ...
Страница 34: ...34 Keysight 1000 Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 88: ...88 Keysight 1000 Series Oscilloscopes User s Guide 3 Capturing Data Figure 37 Alternate Triggers ...
Страница 94: ...94 Keysight 1000 Series Oscilloscopes User s Guide 3 Capturing Data ...
Страница 110: ...110 Keysight 1000 Series Oscilloscopes User s Guide 4 Making Measurements ...
Страница 124: ...124 Keysight 1000 Series Oscilloscopes User s Guide 5 Saving Recalling and Printing Data ...
Страница 155: ......