Specifications and Characteristics
7
Keysight 1000 Series Oscilloscopes User’s Guide
141
Characteristics
NOTE
All characteristics are the typical performance values and are not warranted. Characteristics
are valid after a 30-minute warm-up period and ±5 °C from last calibration temperature.
Table 6
Acquisition System Characteristics
Name
Typical Value
Max sample rate:
2 GSa/s half channel
2
, 1 GSa/s each channel
Memory depth:
20 kpts half channel
2
, 10 kpts each channel
Vertical resolution:
8 bits
Peak detection:
4 ns
Averaging:
Selectable from 2, 4, 8, 16, 32, 64, 128, or 256
Sequence:
Selectable 1 to 800 acquisition frames can be recorded and
played back.
Interpolation:
Sin(x)/x
2
Half channel is when only one channel of channel pair 1-2 is turned on, or one channel of channel
pair 3-4 is turned on.
Table 7
Vertical System Characteristics
Name
Typical Value
Oscilloscope channels:
DSO1xx2A: Ch 1 and 2 simultaneous acquisition
DSO1xx4A: Ch 1, 2, 3, and 4 simultaneous acquisition
Bandwidth (–3dB)
1, 3
:
DSO102xA: DC to 200 MHz
DSO101xA: DC to 100 MHz
DSO100xA: DC to 60 MHz
1
20 MHz when vertical scale is set to < 5 mV (1X probe attenuation).
3
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute
warm-up period and ±10° C from firmware calibration temperature.
Содержание 1000 Series
Страница 12: ...12 Keysight 1000 Series Oscilloscopes User s Guide Contents ...
Страница 16: ...16 Keysight 1000 Series Oscilloscopes User s Guide Tables ...
Страница 34: ...34 Keysight 1000 Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 88: ...88 Keysight 1000 Series Oscilloscopes User s Guide 3 Capturing Data Figure 37 Alternate Triggers ...
Страница 94: ...94 Keysight 1000 Series Oscilloscopes User s Guide 3 Capturing Data ...
Страница 110: ...110 Keysight 1000 Series Oscilloscopes User s Guide 4 Making Measurements ...
Страница 124: ...124 Keysight 1000 Series Oscilloscopes User s Guide 5 Saving Recalling and Printing Data ...
Страница 155: ......