1-
Before Applying Power
Verify that the premises electrical supply is within the range of the instrument.
The instrument has an autoranging power supply.
If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal
condition (in which all means for protection are intact) only.
The Mains wiring and connectors shall be compatible with the connector
used in the premise electrical system. Failure, to ensure adequate earth
grounding by not using the correct components may cause product damage,
and serious injury.
This product is designed for use in Installation Category II and Pollution
Degree 2.
Verify that the premise electrical voltage supply is within the range specified
on the instrument.
When installing the product in a cabinet, the convection into and out of the
product must not be restricted. The ambient temperature (outside the
cabinet) must be less than the maximum operating temperature of the
product by 4 °C for every 100 watts dissipated in the cabinet. If the total
power dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.
This instrument has auto-ranging line voltage input, be sure the supply
voltage is within the specified range.