U3045AM04
Operational Check with Cal Kit
Procedure
1.
Perform an 1-Port Cal of the test set port being verified.
2.
Connect the ECal Module to the Port being verified. Terminate any
remaining ports on the ECal module.
3.
Select the Cal Set to be verified. Select [Cal] > Cal Set... > Cal_File > Apply
Cal > Close. If the "Choose Stimulus Settings..." prompt appears, select
Change the active channel’s stimulus... > OK > Close.
4.
Select Response > Cal > More > ECal > ECal Confidence Check.
5.
In the ECal Confidence Check window, click Change Measurement and
select the test port S-parameter Apply > OK > Read Module Data.
6.
When the "Select ECal Module" prompt appears, select the ECal module
you are using and the ECal Module Memory and Factor ID > OK.
Figure 41
ECal Confidence Check