The Keysight N2140A 200 MHz and N2142A 75 MHz passive 
probes are high impedance probes with switchable attenuation
ratio (1:1 and 10:1). These are compatible with the Keysight 
InfiniiVision x-series oscilloscopes or similar oscilloscopes with
1 Mohm BNC input. 

 Table 3 - Physical & Environmental Characteristics 

Characteristic

Value

Probe Body

Cable Length

Weight

95mm

1200mm

55g

 Table 2 - Electrical Characteristics 

Characteristic 

Value

Attenuation Ratio

Input Capacitance
Compensation range

(when terminated 
into 1 Mohm)

1:1 / 10:1

1:1 - 100pf

15pf - 30pf

Input Resistance

Maximum Input 
Voltage

1:1  - 150V rms CAT II

1:1 - 1MΩ 

Temperature

Humidity

Altitude

The BNC connector of these probes has a 
pin hole without a pin. This is as per the 
design and does not indicate any damage 
or fault in the probe.

These probes have been designed and calibrated for use 

with instruments having an input impedance of 1M Ω 

paralleled by 20pF. However, these may be recalibrated for 
use with instruments having an input capacitance of 
15 to 30pF.

Adjusting Probes for Low-Frequency (LF) Compensation

The probe should be adjusted for LF compensation when it is 
connected to an oscilloscope input for the first time. LF 
compensation matches the probe cable capacitance to the 

oscilloscope input capacitance. This matching assures good 
amplitude accuracy from DC to the upper bandwidth limit 
frequencies. A poorly compensated probe clearly influences 
the overall system performance (probe and oscilloscope) 
and introduces measurement errors resulting in inaccurate 
readings and distorted waveforms. 

10:1 - 15pf

10:1 - 10MΩ

 Table 1 - Probe and its Supplied Accessories 

 ID Part

Quantity

For N2140A

For N2142A

Hook Tip

Ground Spring

Ground Lead

Adjustment Tool

Protective Cap

Channel 
Identification rings

2

2

2

2

-

2

1

1

1

1

 16 rings total 
 (4 rings of each color)

These accessories are provided in the N2141A kit for the 
N2140A probe and the N2143A kit for the N2142A probe.

Undercompensated

Overcompensated

Optimum

10:1  - 300V rms CAT II

3,000 m (9,842 ft) (Operating)
12200 m (40,026 ft) (Non-operating)

Bandwidth

For N2140A
1:1 - 6 MHz
10:1 - 200 MHz

For N2142A
1:1 - 6 MHz
10:1 - 75 MHz

5% to 90% RH up to 30 °C  (Operating)
5% to 65% RH above 30 °C up to 55 °C 
(Non-operating)

-10 °C to +55 °C (Operating)

-40 °C to +70 °C (Non-operating)

Passive probe

2

2

BNC 
Connector

Pin hole 
without 
a pin

      Current Rating 

(DC, 1:1, & maximum voltage)  

0.15mA

To perform LF compensation:

1 Connect the probe to the oscilloscope’s front-panel 
   calibration output (a square wave label is usually seen near 
   this output).
2 Use the supplied adjustment tool to adjust the LF 
   compensation to an optimum square wave response as 
   shown in the figure below.

For optimal probe compensation:

-  Allow the oscilloscope (with the probe connected) to warm
    up for at least 30 minutes.  
-  Repeat the compensation steps if the ambient temperature 
    in the area in which the oscilloscope is being used changes 
    by more than 5 degrees C.
-  Do NOT push down on the adjustment screw when engaging 
    or turning this screw.

To securely engage the hook tip with the 
probe, ensure that you push the hook tip 
onto the probe completely until it stops 
and then release it. 