N9030B PXA Signal Analyzer Service Guide 227
Analog/Digital IF Troubleshooting
A2 Analog IF Assembly Theory of Operation
A2 Analog IF Assembly Theory of Operation
Input Switch and Filter
A 322.5 MHz input signal is received from the A15 Front End Control
Assembly. The signal then goes through a band pass filter centered at
322.5 MHz with a 25 MHz bandwidth. A comb cal signal can be
automatically switched in at the input of the assembly to calibrate the
prefilters.
Step Attenuator and Amplifier
A 1 dB step attenuator follows to compensate for band gain differences.
The signal then goes through a high-dynamic range amplifier.
Image Filters
One of three different filter paths will be selected automatically, depending
on the analyzer mode.
— The through path is selected in IQ Analyzer Mode and wide-band
demod.
— The 12 MHz ceramic bandpass filter is used for normal Signal
Analyzer swept and FFT operation.
— The 300 kHz Surface Acoustic Wave (SAW) is used for ACP modes.
Mixer
There are two mixers that reside on this assembly. The first mixer is used to
down convert the 322.5 MHz input to the final 22.5 MHz IF. The other mixer
is described in the IF Comb Cal section.
3rd LO
The 300 MHz 3rd LO signal comes from the A16 Reference Assembly and
conditions it for use as an LO in the two mixers that reside on this assembly.
The third LO power is ~ +10 dBm to optimize the conversion efficiency of the
first mixer.
IF Comb Cal
The 2nd mixer is used to mix the 300 MHz LO from the Reference Assembly
with the 22.5 MHz cal comb signal from the A3 Digital IF to allow
calibration of the prefilters and overall passband phase and amplitude. The
signal is attenuated by 20 dB if necessary.
.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...