Installation Note N5245-90012 39
Installation Procedure for the Upgrade
3. In the Select Desired Option list, click P04 - 4-Ports.
4. Using the keyboard, enter the license key in the box provided. The license
key is printed on the license message you received from Keysight. Enter
this key exactly as it is printed on the message.
5. Click Enable.
6. Click Yes in answer to the displayed question in the Restart Analyzer? box.
7. Repeat steps 3-6 to enable Option 419, clicking 419 - Src/Rcvr Atten &
Bias Ts 4-Port in step 3.
8. When the installation is complete, click Exit.
Option Verification Procedure
Once the analyzer has restarted and the Network Analyzer program is again
running:
1. On the analyzer’s Help menu, click About Network Analyzer.
2. Verify that “P04” and “419” are listed, but “219” is not listed after
“Options:” in the display. Click OK.
Step 37.Perform Post-Upgrade Adjustments and Calibration
Adjustments
The following adjustments must be made due to the hardware changes of the
analyzer.
— EE default adjustment
— synthesizer bandwidth adj. (only run when EE default adjust is not
sufficient)
— source adjustment
— IF gain adjustment
— receiver characterization
— receiver adjustment
— IF Response Adjustment (Option 090 Only)
— Noise Adjustment (Option 029 Only)
These adjustments are described in the PNA Service Guide and in the PNA
on-line HELP. A list of equipment required to perform these adjustments is also
found in the service guide.
If the options have not been enabled or if the option 219 license has not been
removed, perform the prior steps again. If the options are still not correct, contact
Keysight Technologies. Refer to