72 Installation Note N5222-90128
Installation Procedure for the Upgrade
-
Figure 46
Install the A71–A74 bias-Tee combiner’s gray DC bias cables to the
(N5240-60091)
Route Cables
2.
Route bias cables as shown.
—
Separate cables as much as possible.
—
It is OK to cross the cables.
—
Avoid running cables parallel or next to each other.
—
Avoid tie wrapping to semirigid cables.
J543
J544
J541
J542