background image

C O N F I G U R A T I O N

G U I D E

PNA Family Microwave  

Network Analyzers

(N522x/3x/4xA)

The A-models of the PNA family (N522x/3x/4xA) will be discon-
tinued June 2019. Their B-model replacements are available now. 
For more information about the B-models, please consult the PNA 
“B” configuration guide or www.keysight.com/find/pna

Содержание N5221A

Страница 1: ...Analyzers N522x 3x 4xA The A models of the PNA family N522x 3x 4xA will be discon tinued June 2019 Their B model replacements are available now For more information about the B models please consult the PNA B configuration guide or www keysight com find pna ...

Страница 2: ...onfigurations 21 Application options 22 Upgrade paths 24 Applications 25 Material measurements 25 Signal integrity measurements 25 Multiport measurements 25 Millimeter wave measurements 25 Measurement Accessories 27 Cables and adapter sets 27 Calibration kits 27 For devices with 1 0 mm connectors 28 For devices with 1 85 mm connectors 28 For devices with 2 4 mm connectors 29 For devices with K 2 9...

Страница 3: ... kits and compatible peripherals for the PNA Family microwave network analyzers This guide should be used with the Keysight Technologies Inc PNA Family data sheets for a complete description of these analyzers PNA Series N5221A 10 MHz to 13 5 GHz N5222A 10 MHz to 26 5 GHz N5224A 10 MHz to 43 5 GHz N5225A 10 MHz to 50 GHz N5227A 10 MHz to 67 GHz PNA L Series N5239A 300 kHz to 8 5 GHz N5231A 300 kHz...

Страница 4: ...t adjustment data with measurement uncertainty information compliant to the ANSI NCSL Z540 standard Keysight offers the following options for all PNA Family network analyzers Documentation The PNA Series instruments are equipped with an Online Help system available within the instrument in English only All PNA documentation is available on the web www keysight com find pna Calibration Software Lic...

Страница 5: ...0 kHz to 13 5 GHz N5232A 300 kHz to 20 GHz N5234A 10 MHz to 43 5 GHz N5235A 10 MHz to 50 GHz Description Configurable Test Set Source Attenuators Receiver Attenuators Bias tees Internal Second Source Internal Combiner Mechanical Switches Additional Information PNA X Series Option 200 Option 200 H85 2851 Not available on N5249A or N5247A Option 200 219 Option 200 H85 285 2241 Requires Option 080 No...

Страница 6: ...e measurement uncertainty dynamically real time on the measurement trace Not available with N5231A 32A and 39A Noise figure measurements using standard receivers1 N524xA 028 N522xA 028 n a Requires Option 080 For measuring frequency converters requires Option 082 or 083 Not available with N522xA 200 210 400 and 410 Fully corrected noise figure measurements1 N524xA 029 n a n a Requires Option 080 a...

Страница 7: ...ries N5239A 300 kHz to 8 5 GHz N5231A 300 kHz to 13 5 GHz N5232A 300 kHz to 20 GHz N5234A 10 MHz to 43 5 GHz N5235A 10 MHz to 50 GHz 1 To configure NVNA requires two comb generators with power supplies Keysight calibration kits mechanical or ECal power meter and sensor or USB power sensor Requires EXG MXG or PSG for X parameter extraction PNA X 10 MHz reference output can be used for 10 MHz tone s...

Страница 8: ... N524xA 200 Extended power range and bias tees Option 219 Adds to the standard test set one 65 dB for N5241 42 49A 60 dB for N5244 45A and 50 dB for N5247A source attenuator set table in 5 dB increments for N5241 42 49A and 10 dB incre ments for N5244 45 47A one 35 dB receiver attenuator settable in 5 dB increments for N5241 42 44 45 49A and 50 dB receiver attenuator settable in 10 dB increments f...

Страница 9: ...ther Option H85 485 is not available for the N5249A or the N5247A When ordering select N524xAS then add items N524xAS H85 N524xAS 485 and N524xA 400 Testport3 C R3 Testport1 R1 Testport4 R4 Testport2 R2 A D B Source2 OUT 1 OUT 2 Pulse modulator Source1 OUT 1 OUT 2 Pulse modulator Pulsegenerators To receivers LO R A B C D IF inputs Rearpanel Extended power range and bias tees Option 419 Adds to the...

Страница 10: ...are 60 dB in 10 dB steps and the receiver attenuators are 35 dB in 5 dB steps Also included is a solid state internal RF transfer switch in the R1 reference receiver path This configuration is not available on N5227A 2 ports single source with extended power range and bias tee Option 219 The 2 port test set with extended power range and bias tee comes with a configurable test set source and receiv...

Страница 11: ...attenuators and bias tee at each port For N5221 22A the source attenuators are 65 dB in 5 dB steps and the receiver attenuators are 35 dB in 5 dB steps For N5224 25A the source attenuators are 60 dB in 10 dB steps and the receiver attenuators are 35 dB in 5 dB steps For N5227A both source and receiver attenuators are 50 dB in 10 dB steps Also included is a solid state internal RF transfer switch i...

Страница 12: ... path between a the source output and the reference receiver b the source output and directional coupler thru arm and c the coupled arm of the directional coupler and the port receiver at all ports The source attenuators are 60 dB in 10 dB steps 4 ports single source base configuration Option 400 The 4 port base configuration has no front panel loops and is available only on the N5231A and N5232A ...

Страница 13: ...in source s as well as external signal generators for use as LO signals Supported external sources include the Keysight ESG PSG and MXG Series as well as other SCPI controlled signal generators Option 082 requires Option 080 and cannot be ordered with Option 083 It is compatible with Option 084 which enables measurements of converters with internal LOs Vector and scalar calibrated converter measur...

Страница 14: ...Option 400 or 419 the two internal sources and an unused test port coupler configured as a combiner can be used for two tone IMD measurements When configured with PNA X Option 224 or 423 the two internal sources and internal combiner can be used for two tone IMD measurements Source phase control Option 088 PNA X PNA PNA L This option allows users to set calibrated arbitrary phase differences betwe...

Страница 15: ...PNA X PNA PNA L This option extends the spectrum analyzer Option 090 upper frequency to 110 GHz and beyond when the analyzer is configured with broadband or banded millimeter wave frequency extenders using a millimeter wave controller With this configuration the analyzer s internal receiver attenuators are not in the receiver paths so they cannot be used to reduce signal levels External attenuator...

Страница 16: ...devices Not available with N522xA 200 and 400 Fully corrected noise figure measurements Option 029 PNA X PNA PNA L This option adds a low noise receiver and firmware to the PNA X1 for high accuracy noise figure measurements of amplifiers frequency converters and mixers utilizing Keysight s unique source correction technique Using an Keysight ECal module configured as an impedance tuner for N5241 4...

Страница 17: ...1 OUT 2 Pulse modulator Source 2 OUT 1 OUT 2 Pulse modulator Test port 2 R2 B Noise receivers 10 MHz 6 GHz 6 50 GHz To receivers LO 28V Test port 3 Signal combiner J9 J10 J11 J8 J7 J2 J1 J4 J3 35 dB 35 dB 35 dB 35 dB 60 dB 60 dB 60 dB J6 J5 RF OUT LO OUT R A B C D IF inputs 60 dB Tuner N5241 42 49A 4 port block diagram with noise figure option N5244 45A 4 port block diagram with noise figure optio...

Страница 18: ...in tuner Recommended adapter for f f ECal tuner N4691B 93A 94A 00F 3 5mm85052 60014m madapter N5221 22 41 42 49A 2 4 mm 85056 60005 m m adapter N5224 25 44 45A 1 85 mm 85058 60113 m m adapter N5227 47A 3 5 mm 85052 60014 m m adapter N5241 42 49A N A for N5244 45 47A due to built in tuner Scalar noise calibration available Yes Yes S parameter conversion gain loss measurements Max freq GHz 8 5 13 5 ...

Страница 19: ...re is not available on that series For example PNA Available on PNA Series PNA Not available on PNA Series Pulsed RF measurements Option 008 PNA X PNA PNA L This option adds narrowband pulse detection mode and extends the pulsed RF measurement capability to narrower pulse widths below the limit for the standard wideband detection mode The PNA X and PNA are complete pulsed RF solutions with built i...

Страница 20: ...talled in aircraft Removing switch matrix repeatability errors Removing switch and connector repeatability errors in complex ATE test systems Accessories Solid black series name indicates the feature is available on that series while gray series name with strikethrough indicates the feature is not available on that series For example PNA Available on PNA Series PNA Not available on PNA Series Rack...

Страница 21: ...ed power range to 4 ports N522xA 217 n a N522xAU 617 n a No Expand 2 ports single source with configurable test set extended power range and bias tees to 4 ports N524xA 219 or N522xA 219 N524xAU 942 N522xAU 619 n a No Expand 2 ports internal second source with configurable test set extended power range bias tees combiner and mechanical switches to 4 ports N524xA 224 N524xAU 944 n a n a No Test set...

Страница 22: ... avoid receiver compression for large signal measurements 10 Option 093 is a subset of Option 094 therefore they cannot be ordered together 11 Some parameters are is restricted to 50 GHz Description Required Option For PNA X Series For PNA Series For PNA L Series User Installable Measurement applications Automatic fixture removal Windows 7 OS7 N524xAU 007 N522xAU 007 N523xAU 007 Yes Time domain me...

Страница 23: ... characterization Options 419 and 080 or 400 H85 and 080 N524xAU 5102 n a n a Yes Nonlinear component characterization Export control version Option 419 and 080 or 400 H85 and 080 N524xAU 5112 n a n a Yes Nonlinear X parameters Options 423 and 510 or 511 N524xAU 5143 n a n a Yes Nonlinear pulse envelope domain Requries Options 021 025 and either one of 510 or 511 N524xAU 518 n a n a Yes Arbitrary ...

Страница 24: ...0GHz 67GHz N5221AU 626 N5224AU 650 200 201 217 219 201 217 219 400 401 417 419 401 417 419 600 601 617 619 2 ports 4 ports PNA L upgrade path 200 216 216 400 416 416 Frequency 8 5GHz 13 5 GHz 20GHz 43 5 GHz N5239AU 613 N5231AU 620 50GHz N5234AU 650 2 ports 4 ports N5239AU 620 1 This upgrade path only available on N5241 42 49A 2 Upgrading N5244A units with noise figure hardware requires serial numb...

Страница 25: ...e PNA family of network analyzers to provide network analysis with a single set of connections for devices that have multiple ports The test sets are configured in a variety of ways for extension or switching electro mechanical or solid state switches number of test ports frequency range and 2 or 4 ports depending on your analyzer Test set types switching and extension test sets Switching test set...

Страница 26: ...F access an N5261A 2 port or an N5262A 4 port millimeter wave test set controller with 1 85 mm coaxial cables Option 112 for N5261A Option 114 for N5262A and RF LO IF DC cables sets to select the cable length and quantities and one for 2 ports or two for 4 ports each of right and left frequency extenders with required attenuators and bias tee configurations Waveguide banded solutions to 1 05 THz T...

Страница 27: ...lity for device connection use a cable set Calibration kits Coaxial measurements Mechanical calibration kits include standards such as opens shorts and loads which are measured by the network analyzer for increased measurement accuracy Electronic calibration ECal kits replace mechanical calibra tion standards with one solid state calibration module that is controlled by the network analyzer via US...

Страница 28: ...tor launch assembly For devices with 1 85 mm connectors Mechanical calibration kit 85058B standard DC to 67 GHz Includes 85058 60101 1 85 mm m short 5 4 mm 85058 60102 1 85 mm m short 6 3 mm 85058 60103 1 85 mm m short 7 12 mm 85058 60104 1 85 mm m short 7 6 mm 85058 60105 1 85 mm f short 5 4 mm 85058 60106 1 85 mm f short 6 3 mm 85058 60107 1 85 mm f short 7 12 mm 85058 60108 1 85 mm f short 7 6 ...

Страница 29: ... performance when connected to Keysight Microwave PNA Series network analyzers For devices with 2 4 mm connectors Mechanical calibration kit 85056A standard DC to 50 GHz Includes 00901 60003 2 4 mm m fixed broadband load 00902 60004 2 4 mm f fixed broadband load 00915 60003 2 4 mm m sliding load 00915 60004 2 4 mm f sliding load 85056 60005 2 4 mm m to 2 4 mm m adapter 85056 60006 2 4 mm f to 2 4 ...

Страница 30: ... mm f 11904D 2 4 mm f to 2 92 mm m 11904S 2 4 mm to 2 92 mm Adapter set contains 4 matched adapters 1 Special rugged female connector specifically for connecting to the network analyzer test port but does not mate with a standard male connector For devices with 3 5 mm SMA or 2 92 mm connectors CalPod calibration refresh modules CalPod calibration refresh modules allow in situ calibration refreshes...

Страница 31: ... to 9 GHz 2 ports Standard module includes Option M0F with 85093 60008 3 5 mm f to 3 5 mm m ECal module Option 00F module with 85093 60010 3 5 mm f to 3 5 mm f ECal module Option 00M module with 85093 60009 3 5 mm m to 3 5 mm m ECal module Option 00A adds 85052 60012 3 5 mm f to 3 5 mm f adapter 85052 60014 3 5 mm m to 3 5 mm m adapter 85093C xxx mixed connector options Port A option Port B option...

Страница 32: ...131H1 single flexible 3 5 mm f to 3 5 mm m 62 2 cm 24 5 inches 85134E1 single flexible PSC 3 5 mm f to 2 4 mm f 96 cm 38 inches 85134F1 set flexible One 2 4 mm f to PSC 3 5 mm f 53 cm 21 inches p n 85134 60004 One 2 4 mm f to PSC 3 5 mm m 53 cm 21 inches p n 85134 60003 85134H1 single flexible 2 4 mm f to PSC 3 5 mm m 53 cm 21 inches N4419AK20 single flexible 3 5 mm m to 3 5 mm f 91 4 cm 36 inches...

Страница 33: ...tion 020 module with N4431 60007 4 x Type N f ECal module N4431B xxx mixed connector options Connector type Port A option Port B option Port C option Port D option 3 5 mm f 101 201 301 401 3 5 mm m 102 202 302 402 Type N 50 ohm f 103 203 303 403 Type N 50 ohm m 104 204 304 404 7 16 f 1 105 205 305 405 7 16 m 1 106 206 306 406 N4432A Microwave ECal 300 kHz to 18 GHz 4 ports Includes Option 020 modu...

Страница 34: ...2 4 mm m to WR 22 waveguide adapter For devices with waveguide Mechanical calibration kits X Band X11644A standard WR 90 8 2 to 12 4 GHz Includes 00896 60008 X band standard section 00910 60003 X band termination 11644 20018 X band short 11644 20021 X band shim X281C adapter included in calibration kit WR 90 to 7 mm P Band P11644A standard WR 62 12 4 to 18 GHz Includes 00896 60007 P band standard ...

Страница 35: ... attenuators traceable measured data and uncertainties 85051B 45 MHz to 18 GHz 7 mm kit Includes attenuators airline and mismatch airline with data on a 3 5 inch disk for use in confirming accuracy enhanced system measurement performance traceable to national standards Test procedure is provided in the service manual 85053B 300 kHz to 26 5 GHz 3 5 mm kit Includes attenuators airline and mismatch a...

Страница 36: ...6 5 GHz 83018A power amplifier 2 to 26 5 GHz 27 dB gain to 20 GHz or 23 dB to 26 5 GHz power out 24 dBm to 20 GHz or 21 dBm to 26 5 GHz 83020A power amplifier 2 to 26 5 GHz 30 dB gain to 20 GHz or 27 dB to 26 5 GHz power out 30 dBm to 20 GHz or 26 dBm to 26 5 GHz 83050A power amplifier 2 to 50 GHz 23 dB gain power out 20 dBm to 40 GHz or 17 dBm to 50 GHz 83051A power amplifier 45 MHz to 50 GHz 23 ...

Отзывы: