ENR File Format
5
Keysight N4000/1/2A Operating and Service Manual
55
Viewing the Smart Noise Source ENR Data
To view the ENR data you need a noise figure analyzer and a diskette.
1
Connect the SNS to the NFA Port using the multi-pin 11730 cable, as shown in
.
2
Press the
File
key.
3
Press the
Save
menu key.
4
Press the
ENR
menu key.
5
Press the
SNS
menu key.
6
Press the
Tab
key to highlight the
Path: C:
Ensure ‘
.
’ ‘
^UP^
’ is highlighted.
7
Press the
Select
menu key.
Ensure the ‘
[-A-]
’ directory is highlighted.
8
Press the
Select
menu key.
9
Press the
Enter
key. Wait until the data is saved to the diskette.
10
You can now view the data on a PC using, for example, Wordpad as shown in
The data is presented using Comma Separated Values (CSV). This allows you
to view it in a spreadsheet.
NOTE
For an explanation of saving a file to a diskette using the NFA, see the
Noise
Figure Analyzer User’s Guide
.
NOTE
Ensure the NFA is set to read the SNS data. See
on page 31 for further information.
Содержание N4000A SNS Series
Страница 1: ...Keysight N4000A N4001A and N4002A SNS Series Smart Noise Sources Operating and Service Manual...
Страница 10: ...10 Keysight N4000 1 2A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 12: ...12 Keysight N4000 1 2A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 14: ...14 Keysight N4000 1 2A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 44: ...4 Operating Guide 44 Keysight N4000 1 2A Operating and Service Manual Figure 4 12 Illustrated parts breakdown...