N2893A Probe
N2893A 100 MHz Current Probe User’s Guide
17
Performance Verification for N2893A
Use the following procedure to test the warranted Accuracy specifications for the N2893A
probe (as listed on
page 21
). The recommended test interval to warrant performance of this
probe is once a year. However, you can also test the probe’s performance as and when
required using the recommended test equipment and by following the procedure
documented in this chapter.
Procedure
1
To provide power to the probe, connect the power cord of the N1022B probe
adapter to the Power connector on the front panel of the 1143A power module.
Then connect the N2893A probe’s output to the N1022B probe adapter’s input.
Table 1
Required Test Equipment
Description
Minimum Requirements
Recommended Test Equipment
Oscilloscope
Bandwidth
≥
250 MHz
Amplitude accuracy:
≤
0.4%
Infiniium or InfiniiVision with
1 M
Ω
input
Two Digital Multimeters
-
Keysight U1251B Handheld
Digital Multimeter or equivalent
DVM/DMM
AC Current Generator
AC 10 A
rms
, 50 or 60 Hz, sine
wave
Amplitude accuracy:
≤
0.3%
Keysight AC6800 or Wavetek
9100
Probe Adapter
-
Keysight N1022B Probe Adapter
Power Supply
-
Keysight 1143A Probe Offset
Control and Power Module
Test Cables (with banana jacks) -
-
BNC Cable
-
-
BNC (F) to Banana Adapter
-
-
NMD 3.5mm (M) to 3.5mm (F)
adapter
-
-
3.5mm (M) to BNC (F) adapter
-
-