M9216A Startup Guide

III

WA R N I N G

If this product is not used as spec-
ified, the protection provided by 
the equipment could be impaired. 
This product must be used in a 
normal condition (in which all 
means for protection are intact) 
only.

The types of product users are:

Responsible body is the individual or 
group responsible for the use and main-
tenance of equipment, for ensuring that 
the equipment is operated within its 
specifications and operating limits, and 
for ensuring that operators are ade-
quately trained.

Operators use the product for its 
intended function. They must be trained 
in electrical safety procedures and 
proper use of the instrument. They must 
be protected from electric shock and 
contact with hazardous live circuits.

Maintenance personnel perform rou-
tine procedures on the product to keep 
it operating properly (for example, set-
ting the line voltage or replacing con-
sumable materials). Maintenance 
procedures are described in the user 
documentation. The procedures explic-
itly state if the operator may perform 
them. Otherwise, they should be per-
formed only by service personnel.

Service personnel are trained to work 
on live circuits, perform safe installa-
tions, and repair products. Only properly 
trained service personnel may perform 
installation and service procedures.

Exercise extreme caution when a shock 
hazard is present. Lethal voltage may be 
present on cable connector jacks or test fix-
tures. The American National Standards 
Institute (ANSI) states that a shock hazard 
exists when voltage levels greater than 30V 
RMS, 42.4V peak, or 60VDC are present. 

A good safety practice is to expect that haz-
ardous voltage is present in any unknown 
circuit before measuring.

Operators of this product must be protected 
from electric shock at all times. The respon-
sible body must ensure that operators are 
prevented access and/or insulated from 
every connection point. In some cases, con-
nections must be exposed to potential 
human contact. Product operators in these 
circumstances must be trained to protect 
themselves from the risk of electric shock.

Before operating an instrument, ensure that 
the line cord is connected to a prop-
erly-grounded power receptacle. Inspect the 
connecting cables, test leads, and jumpers 
for possible wear, cracks, or breaks before 
each use.

When installing equipment where access to 
the main power cord is restricted, such as 
rack mounting, a separate main input power 
disconnect device must be provided in 
closed proximity to the equipment and 
within easy reach of the operator.

For maximum safety, do not touch the prod-
uct, test cables, or any other instruments 
while power is applied to the circuit under 
test. ALWAYS remove power from the entire 
test system and discharge any capacitors 
before: connecting or disconnecting cables 
or jumpers, installing or removing switching 
cards, or making internal changes, such as 
installing or removing jumpers.

Do not touch any object that could provide a 
current path to the common side of the cir-
cuit under test or power line (earth) ground. 
Always make measurements with dry hands 
while standing on a dry, insulated surface 
capable of withstanding the voltage being 
measured.

The instrument and accessories must be 
used in accordance with its specifications 
and operating instructions, or the safety of 
the equipment may be impaired.

C A U T I O N

Do not exceed the maximum signal 
levels of the instruments and 
accessories, as defined in the 
specifications and operating 
information, and as shown on the 
instrument or test fixture panels, or 
switching card.

Chassis connections must only be 
used as shield connections for 
measuring circuits, NOT as safety 
earth ground connections.

If you are using a test fixture, keep 
the lid closed while power is 
applied to the device under test. 
Safe operation requires the use of a 
lid interlock.

Instrumentation and accessories 
shall not be connected to humans.

To maintain protection from electric shock 
and fire, replacement components in mains 
circuits - including the power transformer, 
test leads, and input jacks - must be pur-
chased from Keysight. Standard fuses with 
applicable national safety approvals may be 
used if the rating and type are the same. 
Other components that are not 
safety-related may be purchased from other 
suppliers as long as they are equivalent to 
the original component (note that selected 
parts should be purchased only through 
Keysight to maintain accuracy and function-
ality of the product). If you are unsure about 
the applicability of a replacement compo-
nent, call a Keysight office for information.

WA R N I N G

No operator serviceable parts 
inside. Refer servicing to qualified 
personnel.  To prevent electrical 
shock do not remove covers.

Содержание M9216A

Страница 1: ...Keysight M9216A PXI High Voltage Data Acquisition Startup Guide...

Страница 2: ......

Страница 3: ...M9216A Startup Guide I...

Страница 4: ...this document that conflict with these terms the warranty terms in the separate agreement shall control Restricted Rights Legend If software is for use in the performance of a U S Government prime con...

Страница 5: ...possible wear cracks or breaks before each use When installing equipment where access to the main power cord is restricted such as rack mounting a separate main input power disconnect device must be p...

Страница 6: ...U law as of August 13 2005 All electric and electronic equipment are required to be separated from normal waste for disposal Reference WEEE Direc tive 2002 96 EC This symbol indicates the time period...

Страница 7: ...2 Inspect the module for damage 3 Return the module for service 4 Step 2 Verify the Shipment Contents 5 Step 3 Install the Software 6 System requirements 6 Power up the controller 8 Install the softw...

Страница 8: ...VI M9216A Startup Guide...

Страница 9: ...t com find M9216A Related documentation A softcopy of this Startup Guide and the documentation listed below can be found on the CD ROM that is shipped with this product You can also download them for...

Страница 10: ...d at a static safe work station The following figure shows an example of a static safe work station using two types of ESD protection Figure 1 Static safe work station example CAUTION The M9216A is sh...

Страница 11: ...must provide at least 1 of isolation from ground Inspect the module for damage After unpacking the M9216A carefully inspect the unit for any shipping damage Report any damage to the shipping agent im...

Страница 12: ...tag and attach it to the malfunctioning equipment Name and address of owner A P O box is not acceptable as a return address Product model number for example M9216A Product serial number for example M...

Страница 13: ...0001 contains software drivers and all product documentation in PDF format Printed copy of this document M9216A Startup Guide Calibration certificate NOTE Every PXI module is shipped with a Software a...

Страница 14: ...ight IO Libraries Suite 1 5 GB available hard disk space includes 1 GB for Microsoft NET Framework 2 0 SP2 2 65 MB for Keysight IO Libraries Suite 1 5 GB available hard disk space includes 1 GB for Mi...

Страница 15: ...8A chassis Softwares Keysight IO Libraries Suite 16 0 and above is required Adobe Reader version 6 0 or higher is required to view the provided PDF files 1 Because of the installation procedures less...

Страница 16: ...s are located on the bundled CD M9216 10001 The same softwares are also available for free download at the Keysight website http www keysight com find M9216A This installation includes the following K...

Страница 17: ...6A PXI High Voltage Data Acquisition M9216A Startup Guide 9 Figure 2 InstallShield Wizard for Keysight M9216A 3 After installation is complete power down the chassis and the host PC if using the remot...

Страница 18: ...out cooling 4 If you are using an embedded controller proceed to step 5 If you using a remote controller skip to step 6 CAUTION The PXI hardware does not support hot swapping capabilities changing mod...

Страница 19: ...e unlatched downward position Slide the module completely into the chassis When you begin to feel resistance push up on the injector ejector handle to fully inject the module into the chassis 6 If you...

Страница 20: ...sary air circulation in the chassis 11 If you are using a remote controller connect the System Interface Card in the chassis to host computer 12 Plug in and power up the PXI chassis 13 If you are usin...

Страница 21: ...procedures Module installation procedures 1 Press down the ejector 2 Slide the M9216A module into any slot until the backplane connectors touch 3 Seat the M9216A module into the mainframe by pulling u...

Страница 22: ...sight Connection Expert KCE by clicking its desktop shortcut icon or by clicking Start All Programs Keysight IO Library Suite Keysight Connection Expert The KCE will display all modules that are conne...

Страница 23: ...correctly installed in the mainframe If using an external host PC and PCI bus extender system verify that the mainframe power is turned on Ensure that the mainframe is powered on before the PC control...

Страница 24: ...ails on how to use the SFP please refer to the Keysight M9216A Soft Front Panel Help document When successful the M9216A SFP dialog box will appear By default the M9216A will be started in Multiple Ch...

Страница 25: ...e digitizer channel a stable DC source preferably with low noise is required A multimeter can also be used to provide a second verification of the voltage If you intend to verify that the M9216A meets...

Страница 26: ...18 M9216A Startup Guide Keysight M9216A PXI High Voltage Data Acquisition Figure 7 M9216A front panel layout of the input and auxiliary ports...

Страница 27: ...and are able to acquire waveforms You may repeat the steps outlined below to verify all eight digitizer channels in the M9216A For verification purpose the M9216A will be used in Multiple Channel Meas...

Страница 28: ...ted up if not please set the measurement mode to Multiple Channel Note that a hardware reset will occur when switching the measurement mode By default all input ports will be configured with a Record...

Страница 29: ...to the M9216A hardware Note that the relays of the M9216A input port multiplexer require about 1 millisecond to settle After steps A and B the SFP dialog should look similar to Figure 9 In Figure 9 In...

Страница 30: ...age measured with a multimeter Verifying the auxiliary output ports To verify that the M9216A multiplexer is functioning the input ports are switched to the auxiliary output ports The resistance betwe...

Страница 31: ...tiplexer for Bank 1 Table 2 Connection logic of Bank 1 Channel 1 Aux11 Primary Aux12 Complementary Always Connected Port 1 Port 17 Port 9 Port 25 Port 17 Port 1 Port 25 Port 9 Channel 1 Aux21 Primary...

Страница 32: ...e input ports of Channel 2 are able to be switched to the auxiliary output ports To route the input ports to auxiliary output tick the check box Bank 2 Aux Out in the Auxiliary Settings To select a po...

Страница 33: ...e auxiliary output ports Step C Use a multimeter to measure the resistance between the input ports and auxiliary ports For example if input port 18 of Channel 2 was switched to the auxiliary output po...

Страница 34: ...26 M9216A Startup Guide Keysight M9216A PXI High Voltage Data Acquisition...

Страница 35: ...tel 31 20 547 2111 Japan tel 81 426 56 7832 fax 81 426 56 7840 Korea tel 080 769 0800 fax 080 769 0900 Latin America tel 305 269 7500 Taiwan tel 0800 047 866 fax 0800 286 331 Other Asia Pacific Count...

Страница 36: ...This information is subject to change without notice Keysight Technologies 2011 2014 Edition 2 November 2014 M9216 90001 M9216 90001 www keysight com...

Отзывы: