Keysight M9216A PXI High Voltage Data Acquisition
M9216A Startup Guide
21
Step B
Set the DC voltage source to 1.0 V or any suitable voltage.
Connect the DC source to the input port and
channel- under- test either manually or via an external
multiplexer. Select the input port and digitizer channel using
the SFP dialog window. Press
Update Hardware
to ensure that
the input port selection and configuration are updated to the
M9216A hardware. Note that the relays of the M9216A input
port multiplexer require about 1 millisecond to settle.
After steps A and B, the SFP dialog should look similar to
. In Figure 9,
Input Port
10 (
Input Connector
2 of
Channel 2) has been configured and selected for acquisition.
Figure 10
Waveform capture display
N O T E
The SFP uses the
Input Connector
number to select the physical input
port. In this scheme, each of the eight channels has four logical
input
connectors
which is numbered 1 to 4 from top to bottom. Use
help locate the corresponding
input connector
number for any
input port
.
Содержание M9216A