40
1000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Internal Trigger Sensitivity
Generator Setting
Test Limits, greater of
Channel 1 Channel 2
100 MHz models:
10 MHz
0.6 div or 2.5 mV
________
________
Max BW (100 MHz)
1.2 div or 5 mV
________
________
70 MHz models:
10 MHz
0.6 div or 2.5 mV
________
________
Max BW (70 MHz)
0.9 div or 3.8 mV
________
________
50 MHz models:
10 MHz
0.6 div or 2.5 mV
________
________
Max BW (50 MHz)
0.9 div or 3.8 mV
________
________
External Trigger Sensitivity
Generator Setting
Test Limits
Ext Trig In
100 MHz models,
8 V range:
10 MHz
250 mV
________
Max BW (100 MHz)
500 mV
________
100 MHz models,
1.6 V range:
10 MHz
50 mV
________
Max BW (100 MHz)
100 mV
________
70 MHz models,
8 V range:
10 MHz
250 mV
________
Max BW (70 MHz)
500 mV
________
70 MHz models,
1.6 V range:
10 MHz
50 mV
________
Max BW (70 MHz)
100 mV
________
50 MHz models,
8 V range:
10 MHz
250 mV
________
Max BW (50 MHz)
500 mV
________
* Where applicable
Содержание DSOX1102A
Страница 1: ...Keysight InfiniiVision 1000 X Series Oscilloscopes Service Guide ...
Страница 7: ...1000 X Series Oscilloscopes Service Guide 7 Contents Index ...
Страница 8: ...8 1000 X Series Oscilloscopes Service Guide Contents ...
Страница 10: ...10 1000 X Series Oscilloscopes Service Guide Figures ...
Страница 12: ...12 1000 X Series Oscilloscopes Service Guide Tables ...
Страница 52: ...52 1000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 54: ...54 1000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 60: ...60 1000 X Series Oscilloscopes Service Guide Index ...